TC6
®
Repair and Replacement
5-3
6. Position the new latch assembly on the latch/P.C. board panel
and install, but do not tighten, the hardware removed in step 4.
NOTE
Do not loosen the two phillips head screws in center of
latch assembly bracket – these screws secure the latching
mechanism to the bracket, and
must not be loosened.
7. Lower the chamber door until the striker clicks securely into the
latch assembly (the plastic portion of the latching mechanism
will pop up when engaged).
8. Gently lift on the front of the latched door (slotted holes in the
latch assembly bracket will limit upward travel) and position
two 10 mm spacers (for example, the shanks of two 10 mm
screws) between the cabinet door –– one of each side,
approximately two inches from the front corners.
NOTE
When using spacers, care must be taken not to bear
down on the door so that marks are not made on the
cabinet paint or door surface.
9. Gently lower the latched door until it lightly touches the spacers,
then securely tighten the latch assembly mounting hardware.
10. Pull down on the latch solenoid shaft to manually release the
door.
If the latch does not release,
readjust the latch assembly, adjust the
door striker, or (if the door is not parallel to the cabinet) reset the
door hinges. If the latch is still unable to release the door, loosen
the four nuts that secure the latch/P.C. board panel to the frame,
and reset the lateral adjustment of the panel. If none of these
adjustments result in the ability to manually release the door by
pulling down on the latch solenoid shaft, the latch assembly or
striker may need to be replaced.
11. Remove the 10 mm spacers, then adjust and secure the latch
microswitch:
a. Remove the lower mounting screw from the latch
microswitch, apply Loctite
®
222 to the threads, then reinstall
it so that it is just snug (not tight). Do the same with the
upper mounting screw, but leave it loose enough for the
microswitch to pivot on the lower screw.
b. Connect the leads of a DMM (digital multimeter) to the
COM and the N.O. tabs on the microswitch and set the
DMM for ohms.
c. Close the chamber door so that the striker clicks securely
into the latch assembly (the plastic portion of the latching
mechanism will pop up when engaged).
Содержание TC6
Страница 35: ...TC6 Printed Circuit Boards Schematic Diagrams 4 5 4 6 Figure 4 2 System Wiring Diagram SN 9501733 and below ...