Installation and Operation
2-3
POWER
CORD
SELECTED
VOLTAGE
WINDOW
TABS
POWER
CONNECTOR
POWER
CONNECTOR
FUSE
BLOCK
FUSE
BLOCK
VOLTAGE
SELECTOR
VOLTAGE
SELECTOR
VOLTAGE
WINDOW
SELECTABLE
VOLTAGES
FUSES
Figure 2-1. Voltage Selection (SN 9501733 and below)
3. Gently pull the voltage selector from the fuse block.
4. Rotate the voltage selector until the desired voltage is aligned
with the window in the fuse block. Then, reinstall the voltage
selector into the fuse block.
5. Check that proper fuses are installed. Fuses are Type T, 250 V,
but change amperage rating depending on the voltage selected:
•
100 or 110 setting
requires two 3.15-amp fuses, (PN 91428),
•
220 or 240 setting
requires two 2-amp fuses, (PN 91203).
Change fuses if necessary.
6. Reinstall the fuse block into the power connector and plug in
the centrifuge power cord.
NOTE
If the voltage was changed from 100-110 setting to 220-
240 setting, a different power cord is required.
!
C A U T I O N
When changing the voltage
setting, be sure that the voltage
selector is installed so that the voltage
displayed in the voltage window
matches the intended input voltage.
Also, be sure that the proper fuses
have been installed, and that you are
using the proper power cord.
Incorrect installation can result in
damage to the centrifuge. Sorvall is
not responsible for incorrect
installation.
Содержание TC6
Страница 35: ...TC6 Printed Circuit Boards Schematic Diagrams 4 5 4 6 Figure 4 2 System Wiring Diagram SN 9501733 and below ...
Страница 101: ...Illustrated Parts SORVALL Centrifuges 8 2 Figure 8 1 TC6 Main Assembly ...
Страница 103: ...Illustrated Parts SORVALL Centrifuges 8 4 Figure 8 2 TC6 Chassis Assembly ...
Страница 105: ...NOTES ...
Страница 106: ...NOTES ...
Страница 107: ...NOTES ...
Страница 108: ...NOTES ...