Troubleshooting
SORVALL
Centrifuges
6-4
Table 6-1. TC6
®
Troubleshooting Chart (continued)
Problem
Probable Cause
Solution
Door will not open at end of
run. (continued)
Odd readings or missing
READY or DOOR LED's on dis-
play.
Odd reading or missing LED's
on SPEED display.
Rotor turns but no speed dis-
play.
Samples are overheated.
6. Loss of bit in timing string.
7. Loss of tach signal.
1. Loss of segments
2. Open LED.
1. Bad ground to Control
Printed Circuit Board.
1. Loss of a bit, 2 PPS (Pulses
per Second) tach.
2. Internal tach bit delayed.
3. Loss of bit in timing string.
4. Loss of tach signal.
1. Loss of cooling fan power.
2. Cooling fan not working.
1. 30 Vdc supply on Motor Con-
troller Printed Circuit Board
is less than 25 Vdc.
1. Speed set stuck at maximum.
5. Replace DOOR switch.
6. Replace Control Printed
Circuit Board.
7. a. Replace Control Printed
Circuit Board.
b. Replace Motor Control
Printed Circuit Board.
c. Replace motor.
1. Replace the Display Printed
Circuit Board.
2. Replace the Display Printed
Circuit Board.
1. Add a separate ground.
1. Replace Display Printed
Circuit Board.
2. Replace Display Printed
Circuit Board.
3. Replace Control Printed
Circuit Board.
4. a. Replace Control Printed
Circuit Board.
b. Replace Motor Control
Printed Circuit Board.
c. Replace motor.
1. Replace Control Printed
Circuit Board.
2. Replace fan.
1. Locate problem in 30V
supply. Repair.
Содержание TC6
Страница 35: ...TC6 Printed Circuit Boards Schematic Diagrams 4 5 4 6 Figure 4 2 System Wiring Diagram SN 9501733 and below ...
Страница 101: ...Illustrated Parts SORVALL Centrifuges 8 2 Figure 8 1 TC6 Main Assembly ...
Страница 103: ...Illustrated Parts SORVALL Centrifuges 8 4 Figure 8 2 TC6 Chassis Assembly ...
Страница 105: ...NOTES ...
Страница 106: ...NOTES ...
Страница 107: ...NOTES ...
Страница 108: ...NOTES ...