Repair and Replacement
SORVALL
Centrifuges
5-2
5-2. Latch Adjustment
1. Open chamber door.
2. Turn centrifuge power off and unplug power cord.
3. Remove the motor cover and the chamber bowl with door
gasket from the centrifuge.
NOTE
Early model TC6
®
Centrifuges had a ground wire con-
nected to the underside of bowl. The ground wire is no
longer required and should be permanently removed.
4. Remove two phillips head screws securing the bottom of the
latch assembly bracket to the latch/P.C. board panel, then re-
move the two nuts securing the top of the bracket. Refer to
figure 5-1.
5. Disconnect P231, then remove and discard the latch assembly.