S5 System • Fault messages and troubleshooting
SM-5811-0000.01 ENG
5.5
Battery fault
Display in system menu
Description
➜
possible cause/effect
◗
Steps to be taken (please proceed in the order listed)
Battery is a high-drain battery.
The battery charge time was exceeded (internal charge time
limit approx. 3 hours)
Deviation of current measured voltage value from specified
actual value detected internally during trickle charging.
(Internal) fault has occurred during storage of battery capacity.
Battery fault detected.
A software fault has occurred in the UPS.
The following applies to all these messages:
➜
One or both batteries is/are defective
➜
UPS module fault
◗
Replace battery (-ies)
◗
Replace UPS module
Charging time too long
The residual charging time was too long
Fault during trickle charging
No battery capacity stored
Stored battery capacity incorrect
Battery voltage = 0.5 V!
The battery is defective
UPS is defective