S5 System • Fault messages and troubleshooting
SM-5811-0000.01 ENG
5.1
5
Fault messages and troubleshooting
The S5 system undergoes continuous self-monitoring. Faults that occur (e.g. the failure of a component)
are displayed in the system menu.
5.1
Fault messages
5.1.1
Unspecified faults
!
Note
:
Even if
–
all pumps are switched off,
–
the system panel is switched off,
–
the mains power cable is unplugged,
–
and the E/P pack is switched on,
it is still possible that the batteries are totally discharged. In this case, take the appropriate steps
described in chapter 5.1.2.
Fault
Description/possible cause
Steps to be taken
(please proceed in the order listed)
System does not start
although batteries are
charged
Defective switch
– Check/replace switch
System does not start
with discharged or
missing batteries
Mains power cable missing
or defective
Transformer/terminal block
wired wrongly
Defective switch
Defective line filter
Defective rectifier
– Connect/check mains power cable
– Check wiring/terminal block
– Check/replace switch
– Check/replace line filter
– Check/replace rectifier
System starts (can be
heard), but touch screen
remains dark
Circuit breaker for system
panel socket has triggered/
is defective
– Check / reset / replace safety cut-out
Faulty contact
– Check / replace wiring in E/P pack to system panel
socket
– Check / replace system panel cable CAN/24 V
– Check / replace system panel switch
Back plane board of system
panel defective
Replace complete system panel