S5 System • Appendix
SM-5811-0000.01 ENG
7.7
Modules and sensors
Pulse Mode Control for SCP/S5
Pulse frequency
40 to 80 beats per minute
(adjustable in increments of 5)
Pulse flow percentage
50% (not adjustable)
Minimum pulse duration
M
300 ms
Base flow percentage
50% (not adjustable)
Deviation
-15% to +5% (flow rate accuracy between
pulse and non-pulse mode)
APC
Accuracy of volume measurement
(air and blood are suctioned)
when delivering pure liquid ±10% min. ±
20 ml
Accuracy of time measurement
± 1s sec
All sensor modules
Operating voltages
+5 V (from DC/DC module)
24 V (system voltage)
Power consumption
3 W
Level sensor module
Alarm limit (level sensor) for oxygenators/reservoirs made of
rigid polycarbonate, wall thickness at sensor position 3 mm
max.
Level display of the sensor holder ±10 mm
Sensor module for bubble detector and sensor module APC
Alarm limit (bubble sensor) at
N
15 rpm
1/2" and 3/8"
Alarm limit “Large” Air volume: 0.144 cm
3
(Ø 6.5 mm)
Alarm limit “Medium” Air volume: 0.065 cm
3
(Ø 5.0 mm)
Alarm limit “Small” Air volume: 0.034 cm
3
(Ø 4.0 mm)
1/4" and 3/16"
Alarm limit “Small 1/4" ” Air volume:
0.034 cm
3
(Ø 4.0 mm)
Sensor module 2 channel pressure monitor
Accuracy
± 5 mm Hg
Zero point adjustment range
± 100 mm Hg
Gain adjustment range (matching)
± 20%
Input resistance
100 k
T
Output voltage to pressure transducer
< 10 V