S5 System • Appendix
SM-5811-0000.01 ENG
7.9
7.1.3
Pumps
Pump specifications
Roller pump
Double roller pump
Diameter of pump raceway
150 mm
85 mm
Diameter of occlusion roller
30.5 mm
15 mm
Speed range
0 to 250 rpm (clockwise, counterclockwise)
Deviation in speed accuracy
±1% of the terminal value 250 rpm plus ±0.5% of set value
Speed deviation in the event of a fault
(Detection of faulty speed from 30 rpm)
during continuous operation:
+15% max.; 2 revolutions max. until pump stops
Direction of rotation
Clockwise/
counterclockwise
Clockwise/
counterclockwise
Concentricity
Pump raceway
0.03 mm
0.03 mm
Occlusion symmetry
0.03 mm
0.03 mm
Occlusion rollers
0.015 mm
0.015 mm
Displays
rpm display range
0 to 250 rpm
0 to 250 rpm
Resolution
1 rpm
1 rpm
l/min display range (flow)
1/8"
0 to 0.83 l/min
0 to 0.44 l/min
3/16"
0 to 1.79 l/min
0 to 0.93 l/min
1/4"
0 to 3.12 l/min
0 to 1.57 l/min
5/16"
0 to 4.70 l/min
0 to 2.33 l/min
3/8"
0 to 6.50 l/min
o
1/2"
0 to 11.2 l/min
o
Deviation of speed slave pump
max. 1 percentage point of the flow ratio setting
Power supply
Operating voltage
24 V DC
24 V DC
Power consumption
160 W
160 W
Power consumption for mast roller pump
MRP 150: 160 W
MRP 85: 80 W
Device protection
IPX1 (drip-proof)
Touch screen
Pixel Failure Class
Conformity with Pixel Failure Class III