S5 System • Fault messages and troubleshooting
5.4
SM-5811-0000.01
ENG
LED display on pump control panel remains dark.
➜
Connections to circuit board
keys and LEDs (HKD 0327)
interrupted
➜
Fault in circuit board
HKD 0327
➜
(possible) fault in circuit board
HKR 0325
◗
Check connections to
HKD
0327
◗
Replace circuit board
HKR 0327
◗
where necessary replace circuit board
HKR 0325
During maintenance work or when replacing the circuit board,
the DIP switch setting no longer accords with the pump
connections after pump assembly.
➜
The displet of the double roller pump is displayed on
the touch screen of the roller pump (or vice-versa).
➜
Fault in circuit board
HMS 0409
◗
Adjust DIP switch on the
HMS 0409
to the correct
setting:
Switch setting ON: if an RP 150 is connected
Switch setting OFF: if a DRP 85 is connected
◗
Replace circuit board
HMS 0409
An internal malfunction has occured.
◗
Delete message in system menu (twice if necessary).
◗
The pump should not be used. Send pump to the
manufacturer for checking.
Speed knob LED failure
Pump 1
Wrong pump head connected
Pump 1
Short-term internal malfunction,
remains operational- Pump 1
Содержание S5 System
Страница 1: ...S5 System Service Instructions...
Страница 14: ...S5 System Introduction 1 8 SM 5811 0000 01 ENG...
Страница 20: ...S5 System Safety 2 6 SM 5811 0000 01 ENG...