
16
Attaching the effort belts to the patient
Nasal cannula
Attach the nasal cannula to the patient
’s nose. Make sure that the holes of the sensor elements
(highlighted red in the following image) are placed directly below the nostrils.
Note:
Nasal cannulas are single-use items.
Attaching the nasal cannula
SpO2 sensor
The finger sensor should be attached to the most suitable finger. The upper side of the sensor
is marked by the embossing of a finger as shown in the following image. You can use adhesive
tape to secure the cable.
Attaching the finger sensor
Leg EMG sensor
The active electrodes (EMG) should be attached to the skin in the area of the anterior tibialis
muscle of each leg between the knee and ankle (see the following image). Use two adhesive
disposable snap electrodes to place it in the correct position. Connect the electrodes with the
sensor before attaching it to the patient. Use adhesive tape to secure the cable to ensure an
accurate measurement and for strain-relief.
The sensors are labelled L and R and should only be applied to the corresponding leg.
Adjust fixing sleeve
in order to secure the
sensor’s position.