5.0 Setting up the SI7500 (cont.)
502858 issue 1.1
5.0 Setting up the SI7500 (cont.)
24
5.8.4 Nominal With Fixed Limits
A nominal value is displayed between plus and minus fixed limits.
To specify fixed limits above and below a nomi nal value:
1. Press the Limits softkey to specify a nomi nal value with fixed limits.
2. Highlight the desired fields, and then use the numeric keypad to enter val ues.
The example below of 2 mm ± 0.005 mm is used to show a nominal value with fixed limits.
• Nominal = 2.0000 mm
• High warning = 2.0020
• Low warning = 1.9980
• High limit = 2.0040
• Low limit = 1.9960
• Bar graph max = 2.0050