![Solartron Metrology SI7500 Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/solartron-metrology/si7500/si7500_quick-start-manual_1314064022.webp)
5.0 Setting up the SI7500 (cont.)
502858 issue 1.1
5.0 Setting up the SI7500 (cont.)
20
5.8 Limts & Tolerances
The tolerances screen enables the speci fication of nominal values, upper/lower warnings, upper/lower limits and
minimum/maximum bar graph levels for each dimension. Tolerance ranges are indicated numerically and by bands of
color on the Current Value bar and dial screens. The default bar and dial displays show Pass val ues in green, Warning
values in yellow, and Fail values in red.
Cycle through the Setup menu to the Tolerances screen.
Press the right arrow cursor to highlight the Dimension field. Use the increment/decrement softkeys to select the desired
Dimension.
Nominal values with tolerances can be specified as a nominal value b/- tolerances, or as a nominal value with
++/-- tolerances dis tributed entirely on the plus side or entirely on the minus side of the nominal value.