504-2056A 48
Symbol
Description
Caution
Refer to User’s Manual before use
Audio Alarm Silenced/Muted Indicator
This orange light flashes once every 3 seconds during the 60 second silenced
period, to indicate that the electronic alarm is in its silenced state.
It gives a burst of eight flashes every 60 seconds to indicate that the alarm for one
or more functions has been muted.
Low Battery Indicator
This yellow light indicates that battery voltage is low. Initially it flashes every 10
seconds. This indicates that the battery is low although monitoring will continue.
The flashing rate increases to twice every second, accompanied by a medium
priority alarm, for the final few minutes at the end of the battery life.
High Inflation Pressure Visual Alarm
This red light flashes at 2 times per second after the high pressure relief valve has
operated and whenever it is venting inflation gas to the atmosphere. It is
accompanied by a pneumatically generated audible alarm. If the relief valve has
operated for 3 successive cycles, or for at least one half second, then the high
priority audible alarm sounds and the visual alarm flashes continuously. Both
persist until 10 seconds after the condition is last detected. They then automatically
reset.
The same alarm signals are given if a constant positive pressure greater than 10
x100Pa is detected for 8 seconds or more.
If the constant positive pressure alarm has been muted then the flashing of this
alarm decreases to a burst of 3 flashes every 8 seconds.
Cycle Indicator
This green light flashes once, for 1/10
th
second, every time the patient inflation
pressure rises through the pre-set threshold pressure of 10 x100Pa (10cmH
2
O).
This indicates normal operation.
Low Pressure/Disconnect Visual Alarm
This yellow light flashes 30 times per minute if 'cycle detect' or 'breathing detect'
has not been activated for 10 seconds. It is accompanied by a medium priority
alarm.
Single Gas Operations
This green light gives a burst of 3 flashes every 30 seconds whenever the ventilator
is operating on a single gas supply (oxygen or air only). If one of the gases fails and
audible alarm will sound until muted (section 2 (c) item11).
Y
Collect Separately
This product contains electronic and other components (such as batteries) that may
contain materials which, if disposed of with general household waste, could be
damaging to the environment.
In accordance with Directive 2002/96/EC Waste Electrical and electronic
Equipment, Smiths Medical International Ltd. requires that residents of the
European Union return this product for proper disposal at the end of its useful life.
Contact your local distributor for specific instructions on how to return the product
for disposal.
Содержание Pneupac babyPAC 100
Страница 2: ...504 2056A 2 Page Intentionally Blank ...
Страница 10: ...504 2056A 10 Page Intentionally Blank ...
Страница 26: ...504 2056A 26 Page Intentionally Blank ...
Страница 38: ...504 2056A 38 Page Intentionally Blank ...
Страница 41: ...504 2056A 41 Figure 4 ...
Страница 42: ...504 2056A 42 Page Intentionally Blank ...
Страница 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Страница 52: ...504 2056A 52 Page Intentionally Blank ...
Страница 56: ...504 2056A 56 Page Intentionally Blank ...
Страница 57: ...504 2056A 57 APPENDIX B Calibration accuracies and deviations due to change in ambient conditions ...
Страница 58: ...504 2056A 58 Page Intentionally Blank ...
Страница 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...
Страница 62: ...504 2056A 62 Page Intentionally Blank ...
Страница 64: ...504 2056A 64 Page Intentionally Blank ...