504-2056A 23
SECTION 3
: SET-UP AND FUNCTIONAL CHECK
(a) Set
Up
Unpack the babyPAC ventilator control module and accessories and check all items against
the contents checklist. If any items are missing or incorrect, or have become damaged,
notify your supplier immediately.
WARNING:
To avoid harm to the patient, user’s should be aware that it is important to check
the patient circuit connectors for damage, as damaged connectors could adversely
effect ventilator performance.
If a carrying sling or mounting bracket has been supplied attach it to the module as shown in
Figure 2. Insert the battery into the module battery holder as described in Section 6(c).
Because there is no battery fitted in the holder it will be necessary to tip up the module in
order to remove the battery retaining cap. Connect the patient circuit to the control module
as shown in Fig 1 (a). Connect the input hose(s) to the appropriate inlet connector(s) as
shown in Figure 1(a) and tighten the securing nut(s) lightly using a spanner (wrench).
The probe(s) on the input hose(s) is gas specific to the standard specified when ordering.
Any of the Pneupac
®
hoses listed in the babyPAC Accessories and Spare Parts List can be
used with the ventilator provided the same inlet connection is specified.
(b) Functional
Check
The following procedure should be followed when first setting up the ventilator to check
that it has been assembled correctly and is operating safely. It should be repeated
periodically as specified under 'Maintenance'.
1.
Set the ventilator controls as follows:-
Function
Selection
Switch:
0
Inspiratory
Time:
1.0
second
Expiratory
Time:
2.0
seconds
Inspiratory
Pressure:
20
x100Pa
(20
cmH
2
O)
Audible High Pressure Alarm:
30 x100Pa (30 cmH
2
O)
PEEP Pressure:
0
Oxygen Concentration:
< 50% (white scale)
2.
Connect the probe(s) on the input hose(s) to an appropriate gas outlet(s).
WARNING:
Oil, grease or combustible lubricants, other than those approved for oxygen service,
must never be allowed to come into contact with the parts of the ventilator, oxygen
regulator or cylinder. Particular care should be taken to avoid any trace of
contamination around the oxygen inlet and outlet ports. Oil or Grease readily
oxidise and in the presence of oxygen will violently burn. Avoid smoking or naked
flame.
WARNING:
Always connect the ventilator to the regulator before opening the cylinder valve and
then open the valve slowly in order to avoid the risk of ignition induced by the heat
generated by adiabatic compression.
Содержание Pneupac babyPAC 100
Страница 2: ...504 2056A 2 Page Intentionally Blank ...
Страница 10: ...504 2056A 10 Page Intentionally Blank ...
Страница 26: ...504 2056A 26 Page Intentionally Blank ...
Страница 38: ...504 2056A 38 Page Intentionally Blank ...
Страница 41: ...504 2056A 41 Figure 4 ...
Страница 42: ...504 2056A 42 Page Intentionally Blank ...
Страница 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Страница 52: ...504 2056A 52 Page Intentionally Blank ...
Страница 56: ...504 2056A 56 Page Intentionally Blank ...
Страница 57: ...504 2056A 57 APPENDIX B Calibration accuracies and deviations due to change in ambient conditions ...
Страница 58: ...504 2056A 58 Page Intentionally Blank ...
Страница 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...
Страница 62: ...504 2056A 62 Page Intentionally Blank ...
Страница 64: ...504 2056A 64 Page Intentionally Blank ...