504-2056A 30
is not generally recommended because the increased compliance will affect the volume which can
be delivered by the ventilator for larger patients and the maximum effective frequency for smaller
babies. It is also more susceptible to kinking and inadvertent flattening.)
All parts which contain metals and which are MR Compatible are marked "/MRI" on their
packaging and where this feature is required for additional parts and accessories /MRI should be
added as a suffix to the order code of each part ordered (See Section 7).
WARNING:
The normal routine of checking the ventilator system for magnetic attraction should
be followed whenever the equipment is taken into an MR environment to ensure
that magnetic parts have not been added to the system inadvertently. It is also
recommended that every time a patient is positioned in the magnetic field the
manometer is checked to confirm unchanged ventilation and the high pressure
relief/alarm system is checked by a temporary circuit disconnection and occlusion of
the ventilator outlet connector.
Содержание Pneupac babyPAC 100
Страница 2: ...504 2056A 2 Page Intentionally Blank ...
Страница 10: ...504 2056A 10 Page Intentionally Blank ...
Страница 26: ...504 2056A 26 Page Intentionally Blank ...
Страница 38: ...504 2056A 38 Page Intentionally Blank ...
Страница 41: ...504 2056A 41 Figure 4 ...
Страница 42: ...504 2056A 42 Page Intentionally Blank ...
Страница 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Страница 52: ...504 2056A 52 Page Intentionally Blank ...
Страница 56: ...504 2056A 56 Page Intentionally Blank ...
Страница 57: ...504 2056A 57 APPENDIX B Calibration accuracies and deviations due to change in ambient conditions ...
Страница 58: ...504 2056A 58 Page Intentionally Blank ...
Страница 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...
Страница 62: ...504 2056A 62 Page Intentionally Blank ...
Страница 64: ...504 2056A 64 Page Intentionally Blank ...