504-2056A 8
If only oxygen is available as a compressed supply gas then an oxygen
concentration is
available within the following range:-
NOTE: The pointer on the control knob rotates beyond the 50% marker.
•
Control set to minimum setting (fully anti-clockwise), approximately 45%
concentration.
•
Control set to maximum setting (fully clockwise) 100% concentration.
NOTE:
Oxygen concentrations
– The user should be aware that the oxygen concentration could
be as low as 93%, when the ventilator is set at a high ventilation frequency and the 100%
O
2
position.
17.
Warning: Damaged Patient Circuit Connectors (Section 3 (a))
To avoid harm to the patient, user’s should be aware that it is important to check the patient
circuit connectors for damage, as damaged connectors could adversely effect ventilator
performance.
18.
WARNING: Heat Moisture Exchanger (HME) filter (Section 4 c))
If a Heat Moisture Exchanger (HME) filter is used and the patient is experiencing breathing
difficulties, or there are mechanical ventilation problems, or decreases in gas exchange, always ensure
that there are no blockages in the breathin system. If a blockage is found, immediately replace the
breathing system or HME filter. To prevent such problems occuring, filters and HME’s shiould be
routinely replaced in accordance with the manufacturer’s instructions.
19.
WARNING: Exhalation Valve Diaphragm Fitment (Section 5 f))
To avoid the risk of the exhalation valve diaphragm remaining
depressed against the valve housing and resulting in continuous
pressure loss from the patient circuit, avoid depressing the center of
the diaphragm during fitment. Following the replacement procedure,
ensure that the center of the diaphragm is is not depressed and always
perform a functional check.
PRECAUTIONS: -
1.
Protection of Battery and Ventilator During Periods Without Use (Sect. 2(c)15)
To avoid any drain on the battery, if the ventilator is unlikely to be used for a long time or is placed
in storage, ensure that the function selector switch is left in the ‘O’ (ventilator off) position. It is
also recommended that the battery is removed from its holder to avoid possible corrosion due to
leakage of its contents.
Содержание Pneupac babyPAC 100
Страница 2: ...504 2056A 2 Page Intentionally Blank ...
Страница 10: ...504 2056A 10 Page Intentionally Blank ...
Страница 26: ...504 2056A 26 Page Intentionally Blank ...
Страница 38: ...504 2056A 38 Page Intentionally Blank ...
Страница 41: ...504 2056A 41 Figure 4 ...
Страница 42: ...504 2056A 42 Page Intentionally Blank ...
Страница 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Страница 52: ...504 2056A 52 Page Intentionally Blank ...
Страница 56: ...504 2056A 56 Page Intentionally Blank ...
Страница 57: ...504 2056A 57 APPENDIX B Calibration accuracies and deviations due to change in ambient conditions ...
Страница 58: ...504 2056A 58 Page Intentionally Blank ...
Страница 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...
Страница 62: ...504 2056A 62 Page Intentionally Blank ...
Страница 64: ...504 2056A 64 Page Intentionally Blank ...