504-2056A 28
12.
Temporarily disconnect the ventilator circuit at the connection closest to the patient
and verify that the green cycle light fails to illuminate during an inspiratory phase.
Immediately reconnect circuit.
The patient’s condition and chest movements as well as the Inflation Pressure Monitor
should be kept under constant observation so that adverse ventilation conditions can be
detected and corrected before the patient is put at risk.
WARNING:
At all times check adequacy of gas supply by observing the gas failure alarm and the
gas cylinder contents indicator.
Table 1
(c) Humidification
There is no provision in the standard babyPAC patient circuit for a heated water humidifier because
of the problems these present during the transportation of patients. If humidification is considered
necessary then the use of alternative means such as HMEs is recommended. These should comply
with EN ISO 9360.
WARNING: If a Heat Moisture Exchanger (HME) filter is used and the patient is experiencing
breathing difficulties, or there are mechanical ventilation problems, or decreases in gas
exchange, always ensure that there are no blockages in the breathin system. If a blockage
is found, immediately replace the breathing system or HME filter. To prevent such
problems occuring, filters and HME’s shiould be routinely replaced in accordance with
the manufacturer’s instructions.
Содержание Pneupac babyPAC 100
Страница 2: ...504 2056A 2 Page Intentionally Blank ...
Страница 10: ...504 2056A 10 Page Intentionally Blank ...
Страница 26: ...504 2056A 26 Page Intentionally Blank ...
Страница 38: ...504 2056A 38 Page Intentionally Blank ...
Страница 41: ...504 2056A 41 Figure 4 ...
Страница 42: ...504 2056A 42 Page Intentionally Blank ...
Страница 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Страница 52: ...504 2056A 52 Page Intentionally Blank ...
Страница 56: ...504 2056A 56 Page Intentionally Blank ...
Страница 57: ...504 2056A 57 APPENDIX B Calibration accuracies and deviations due to change in ambient conditions ...
Страница 58: ...504 2056A 58 Page Intentionally Blank ...
Страница 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...
Страница 62: ...504 2056A 62 Page Intentionally Blank ...
Страница 64: ...504 2056A 64 Page Intentionally Blank ...