504-2056A 25
13.
Finally, set the controls as specified in step 1 so that the ventilator is left set for emergency
use.
WARNING:
Any deviations observed in the above checking should be reported immediately to
your supplier, or an authorised service engineer, and the ventilator should not be
used until a more thorough check has been completed.
WARNING:
Before changing gas cylinders, turn off the cylinder valve and then switch on the
ventilator. After one or two cycles the ventilator will stop and it is then safe to
unclamp the pin index yoke without a sudden release of pressure.
PRECAUTION:
It is recommended that the valve on the gas cylinder is turned off after use to
ensure that the cylinder contents are not lost during storage due to small leakages.
Содержание Pneupac babyPAC 100
Страница 2: ...504 2056A 2 Page Intentionally Blank ...
Страница 10: ...504 2056A 10 Page Intentionally Blank ...
Страница 26: ...504 2056A 26 Page Intentionally Blank ...
Страница 38: ...504 2056A 38 Page Intentionally Blank ...
Страница 41: ...504 2056A 41 Figure 4 ...
Страница 42: ...504 2056A 42 Page Intentionally Blank ...
Страница 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Страница 52: ...504 2056A 52 Page Intentionally Blank ...
Страница 56: ...504 2056A 56 Page Intentionally Blank ...
Страница 57: ...504 2056A 57 APPENDIX B Calibration accuracies and deviations due to change in ambient conditions ...
Страница 58: ...504 2056A 58 Page Intentionally Blank ...
Страница 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...
Страница 62: ...504 2056A 62 Page Intentionally Blank ...
Страница 64: ...504 2056A 64 Page Intentionally Blank ...