Level 1®
CONVECTIVE WARMER
| Operator’s Manual
29
S E C T I O N 8 • S p e c i f i c a t i o n s
S E C T I O N 8
Specifications
System Design Specifications
Compliance
Product Safety
Convective Warming Blanket Flammability
EMC
Enclosure Protection
Convective Warmers
Drop Test Compliance
Vibration Test Compliance
Standard
IEC 60601-1 Ed. 3.1
NFPA 702, 16 CFR 1610
IEC 60601-1-2 Ed. 3.0 and 4.0, FCC 47 CFR Part 15, Class B
IEC 60529 Ed. 2.2, IP Code: IP12
IEC 80601-2-35 Ed. 2.0 and 2.1
IEC 60601-1 Ed. 3.1
IEC 60068-2-64 Ed. 2.0
Physical
Height, Overall
Width, Overall
Depth, Overall
Weight, Overall
Weight, Max. Working Load
Hose, Length
Power Cord, Length
Maximum Air Flow
Filtration System Media
Tip Test (Cart Mounted)
Shipping Durability
Dimensions
32 cm
39 cm
28 cm
7.2 kg
10.2 kg
2 m
3 m, 5 m
48 cfm at 0.4 inches of H
2
0 backpressure
99.97% or greater for 0.2
m
m size particles.
BFE & VFE efficiency 99.999% tested per
ASTM F2101
Meets IEC 60601-1 and EN 60601-1
Meets ASTM D4169 - Dist. Cycle 12, Assurance Level I.
Vehicle stack height meets Assurance Level II.
Environmental
Operation
Operation Altitude
Transportation & Storage
Transportation & Storage Pressure
Temperature
Humidity
10°C to 40°C 15% to 95% noncondensing
-400 to 3000 meters
-40°C to 60°C 15% to 95% noncondensing
510 mm Hg (67.83 kPa) to 795 mm Hg (105.74 kPa)
Thermal
Hose End Temperature
Maximum Contact Surface Temperature
Hose End Temperature Control
The Snuggle Warm® convective warming
blankets meet the average and maximum contact
surface temperature
Response time to set point (with the
lowest back pressure specification of 0.4 inches H
2
O)
Temperature
Settings: Ambient, 37°C, 40°C and 43°C
Distal temperature within ± 1°C of setting
44°C
IEC: 80601-2-35
< 120 seconds at 120 VAC, at 18⁰C ambient room temperature
< 120 seconds at 230 VAC, at 18⁰C ambient room temperature
< 120 seconds at 100 VAC, at 20⁰C ambient room temperature