Only use accessories that are specified for use
with the CADD
®
-Solis ambulatory infusion
pump.
•
If you are using a medication cassette reservoir
in which the medication is frozen, thaw at
room temperature only. Do not heat in a
microwave oven as this may damage the
product and cause leakage.
•
If the power up results in an error message
indicating that the protocol library was lost, do
not proceed with using the pump. Follow your
facility’s procedures for downloading protocol
libraries.
•
Do not immerse the pump in cleaning fluid
or water. Do not allow solution to soak in to
the pump, accumulate on the keypad, or enter
the battery compartment, USB port, remote
dose cord jack, or power jack areas. Moisture
buildup inside the pump may damage the
pump.
•
Do not clean the pump with acetone, other
plastic solvents, or abrasive cleaners, as
damage to the pump may occur.
•
The pump SHOULD NOT BE DIRECTLY
IRRADIATED by therapeutic levels of ionizing
radiation because of the risk of permanent
damage to the pump’s electronic circuitry.
The best procedure to follow is to remove the
pump from the patient during therapeutic
radiation sessions or diagnostic levels of
radiographic and fluoroscopic radiation. If
the pump must remain in the vicinity during
a diagnostic or therapy session, it should be
shielded, and its ability to function properly
should be confirmed following treatment.
•
Do not expose the pump directly to
ultrasound, as permanent damage to the
pump’s electronic circuitry may occur.
•
Magnetic fields produced by magnetic
resonance imaging (MRI) equipment may
adversely affect the operation of the pump.
Remove the pump from the patient during
MRI procedures and keep it at a safe distance
from magnetic energy.
•
Use of this pump on patients monitored by
electronic equipment may cause artifactual
interference. As with all electronic equipment,
electrical artifacts which affect the
performance of other equipment, such as ECG
monitors, can occur. The user should check
the correct function of the equipment prior to
use.
•
Do not use the pump in hyperbaric chambers
as they will affect how the pump works and
may also cause damage to the pump.