27
Programming and Operation
•
PCA Doses Attempted
shows the total
number of PCA doses attempted by the patient
while the pump was running, including those
that were delivered, locked out, and stopped in
progress.
The dose counters can be viewed or cleared while
the pump is running or stopped.
NOTE
: If the PCA dose is not available in the
current protocol and has been programmed to not
be visible on the home screen, you will only see the
“Total Given” when viewing this report.
To view the given and PCA dose counters:
1. From the reports menu,
press
U
or
D
to
choose
Given and PCA
Dose Counters
and
press
S
.
To clear the total given:
1. Press
U
or
D
until
the Total Given section
is highlighted.
2. Select
Clear Given
.
To clear the PCA doses given and attempted:
1. Press
U
or
D
until
the PCA Doses Given
and Attempted section
is highlighted.
2. Select
Clear Doses
.
Adjust Admin Settings
The administrator settings contain pump
configurations that are set up by the CADD®-Solis
system administrator. Protocol libraries are created
using the CADD®-Solis Medication Safety Software.
The administrator settings only allow you to make
changes to the protocol currently displayed.
To access the administrator settings:
1. From the tasks menu,
Press
U
or
D
to
choose
Adjust Admin
Settings
and press
S
.
NOTE
: You cannot
access the administrator
settings without the administrator code.
See the Administrator Settings Guide for more
information.
Reports
The reports screen is used to access a variety of
reporting and record-keeping functions.
You may access the reports
from the home screen or
the tasks menu. To access
the reports from the home
screen, select
Reports
.
(To access the reports from
the tasks menu, see “View
Reports” in the previous
section.)
Given and PCA Dose Counters
This screen shows the number of PCA doses given
and attempted since the date and time indicated,
which is the last time they were cleared manually, or
when a new protocol or new patient has been started.
•
Total Given
shows the amount of drug (in
programming units) that has been given in
continuous rate, clinician boluses, and PCA
doses.
•
PCA Doses Given
shows the number of
PCA doses actually delivered to the patient,
including any doses stopped in progress.
Press here to select
Press here to select
Press here to select