46
References and Troubleshooting
EN 55011 (1998),
Industrial, scientific and
medical (ISM) radio frequency equipment
- Radio disturbance characteristics - Limits
and methods of measurement. Amendment 1
(1999), Amendment 2 (2002)
Miscellaneous Standards
USB 1.1
Universal Serial Bus (USB)
Specification, Revision 1.1, September 23,
1998 – USB.org
EN 1041 (1998),
Information supplied by the
manufacturer with medical devices.
IEC/TR 60878 (2003)
, Graphical symbols for
electrical equipment in medical practice.
EN 980 (2003)
, Graphical symbols for use in
the labeling of medical devices.
IEC 60529 (2001)
, Degrees of protection
provided by enclosures (IP Code).
Specifications (Nominal)
General Pump Specifications
Medication cassette reservoir, part number 21-7002,
CADD
®
extension sets, part number 21-7045, and
CADD
®
administration sets, part number 21-7091
were used to test the pump.
Resolution
Medication cassette reservoir or CADD
®
administration set, 0.050 mL per pump stroke
nominal.
Size
4.1 cm × 10.2 cm × 12.7 cm [1.6 in. × 4 in. ×
5 in.] excluding cassette or other accessories.
Weight
595 g [21 oz.] including 4 AA alkaline batteries,
excluding other accessories.
Pump Alarms
The following alarms are all considered to be
high priority alarms: Battery depleted while
delivering; Battery removed while delivering;
Battery unusable while delivering; Rechargeable
battery end of life; Disposable detached while
delivering; Disposable attached improperly;
Disposable type high flow administration set;
Disposable type invalid; Disposable locked but
not latched; Reservoir volume empty; Pressure
sensor faulty; Downstream occlusion; Upstream
occlusion; Air in line detected; Remote dose
cord key stuck; Key stuck; Stop mode reminder;
Pump was automatically stopped.
There are an additional 19 medium
priority alarms, 8 low priority alarms, and
20 informational messages alerts.
Battery Fallout Alarm
Alarm sounds for 2 minutes if the pump has
been powered up for a minimum of 2 minutes.
NOTE
: Only enabled while pump is in run
mode.
Battery Status
Battery State
CADD® Pump Status
25%—100%
No alarm
Low battery
Transition to low battery
condition; battery low message
appears; 3 beeps every 5 min.
†
LCD backlight will flash for 12 ms
during each motor operation.
Depleted
battery
Transition to depleted battery
condition; battery depleted
message appears; pump beeps
continuously.
††
Shut down
Pump shuts off due to too low
operating voltage.
CADD®-Solis pump low battery conditions.
†
The pump emits 3 beeps every 5 minutes, and the low battery warning
message appears on the pump’s display, indicating that the battery
power is low, but the pump is operable.
††
The pump emits a continuous, variable-tone alarm, and the depleted
battery warning message appears on the display, the battery power is
too low to operate the pump, and the pump delivery operation has
stopped.
Continuous Delivery Alkaline Battery Life
with screen intensity set to 3
Rate
Life
Volume
0.4 mL/hr
260 hrs
104 mL
1 mL/hr
210 hrs
210 mL
5 mL/hr
165 hrs
825 mL
10 mL/hr
150 hrs
1500 mL
30 mL/hr
70 hrs
2100 mL
Содержание CADD-Solis