33
Programming and Operation
The following chart includes the initial settings of
the Manual Mode.
Pump Function
What Changes in the Protocol
Programming Units
User chooses mL, mg, or mcg
Concentration
Range is 0.1–100 mg or/mL or
1–500 mcg/mL
Continuous Rate
Range is 0–30 mL/hour (or mg or
mcg equivalent), Default 0
PCA dose
Range is 0–20 mL (or mg or mcg
equivalent), Default 0
PCA dose lockout
Range is 1 min–24 hours,
Default 1 hr (when PCA dose is
programmed).
Delivery limit amount* Range is 0.1–1000 mL (or mg or
mcg equivalent)
Max doses/hour*
Range is 1–60
Reservoir volume
1 mL
Reservoir volume reset
value
100 mL
Delivery limit method
Remains the same as the
previously used protocol
Reservoir volume low
trip point
5.0 mL
Reservoir low alarm
type
Remains the same as the
previously used protocol
Clinician bolus amount Range is 0–20 mL (or mg or mcg
equivalent)
Delivery limit period
Remains the same as the
previously used protocol
Maximum delivery rate Remains the same as the
previously used protocol
Pump stopped alarm
type
Remains the same as the
previously used protocol
Upstream sensor on/off Remains the same as the
previously used protocol
Downstream sensor
sensitivity
Remains the same as the
previously used protocol
Air detector on/off
Remains the same as the
previously used protocol
Air detector sensitivity
Remains the same as the
previously used protocol
Time and date settings
Remains the same as the
previously used protocol
PM reminder on/off
Remains the same as the
previously used protocol
PM reminder interval
Remains the same as the
previously used protocol
Display and sound
settings
Remains the same as the
previously used protocol
Security settings
Remains the same as the
previously used protocol
*Only if feature was used in the previous protocol.
To use the manual mode:
1. Determine whether you are starting a new
patient, or starting a new protocol with the same
patient. Select the appropriate task from the
tasks menu.
2. On the Select Therapy
screen, scroll to the
bottom of the therapies
until [Program
Manually] is highlighted
and press
S
.
The protocol title bar will be displayed in the
color black. Instead of a therapy and qualifier, you
will see the Manual Program screen (see page 32).
NOTE
: You may need to enter a higher level
security code to proceed. If you do not know
which code to enter, contact your CADD®-Solis
system administrator.
3. Press
U
or
D
to
select the desired units
(mL, mg, mcg) and
press
S
.
If you selected mg or
mcg, press
U
or
D
to select the desired
concentration and select
Confirm
. If you selected
mL, proceed to the next
step.
4. The therapy,
qualifier, units and
concentration, if
applicable, that you
have selected will
now appear. Confirm
that you have selected
the correct units and
concentration and
select
Yes
.
Press here to Confirm
Press here to select