39
References and Troubleshooting
Alarms and Messages
Alarm Priority
Description / Corrective Action
Delivery limit reached. (Pump’s
status bar will read “KVO = 0”)
OR
Delivery limit reached and
partial dose delivered. (Pump’s
status bar will read “KVO = 0”)
Low
The programmed delivery limit has been reached, and the
pump is not delivering fluid. This alarm occurs when the
continuous rate is programmed to 0 mL/hr, and either a PCA
dose or the continuous rate has caused the delivery limit to be
exceeded. Select acknowledge to clear the alarm or the alarm
will automatically clear after 5 seconds.
NOTE
: The status bar display will remain after the alarm has
cleared.
Delivery limit reached.
(Pump’s status bar will read
“Del Limit”)
OR
Delivery limit reached
and partial dose delivered.
(Pump’s status bar will read
“Del Limit”)
Low
The programmed delivery limit has been reached, and the
pump is delivering fluid at the KVO rate of 0.1 mL/hr. This
alarm occurs when the continuous rate is programmed to
greater than 0 mL/hr, and either a PCA dose or the continuous
rate has caused the delivery limit to be exceeded. Select
acknowledge to clear the alarm or the alarm will automatically
clear after 5 seconds.
NOTE
:
The status bar display will remain after the alarm has
cleared.
Delivery too slow. 1 mL of fluid
not delivered.
Medium
The pump is busy with too many activities and does not have
sufficient resources to support the programmed delivery rate.
The delivery has fallen behind by 1 mL. Select acknowledge
to clear the alarm. If this alarm occurs regularly, the pump
may be faulty. Remove the pump from use and contact Smiths
Medical Customer Service.
Depleted battery is charging.
Low
The rechargeable battery pack is depleted and is being
recharged with the AC adapter. Select acknowledge to clear the
alarm or the alarm will automatically clear after 5 seconds.
Downstream occlusion. Clear
occlusion between pump and
patient.
High
The pump has detected high pressure, which may be resulting
from a downstream blockage, kink in the fluid path, or a
closed tubing clamp. Delivery is paused and will resume if
the occlusion is removed. Remove the obstruction to resume
operation. Or select
Stop Pump
to stop the pump and silence
the alarm for 2 minutes, then remove the obstruction and
restart the pump.
NOTE
: To reduce the potential bolus delivery after an
occlusion, perform the following:
1. Select
Stop Pump
to stop the pump.
2. Close the distal clamp. If the distal clamp is the cause of
obstruction, keep it closed and continue with step 4.
3. Remove the obstruction.
4. Detach the medication cassette reservoir or CADD®
administration set from the pump.
5. Open the flow stop feature, if present.
6. Wait 10 seconds.
7. Close the flow stop feature, if present.
8. Reattach the medication cassette reservoir or CADD®
administration set to the pump.
9. Open the distal clamp.
10. Review the pump’s program.
11. Restart the pump.