37
References and Troubleshooting
Alarms and Messages
Alarm Priority
Description / Corrective Action
Battery removed. Pump
stopped.
High
The rechargeable battery pack or the 4 AA batteries were
removed. The pump was delivering and is now stopped. Select
acknowledge to clear the alarm.
• Install a fully charged rechargeable battery pack or 4 new
AA batteries. In order to start delivery, good batteries must
always be installed, even when an external source of power is
connected.
• If appropriate, restart the pump.
Battery removed. Pump will
not run.
Medium
The pump is stopped and the rechargeable battery pack or the
4 AA batteries were removed, but the pump is still powered by
the AC adapter. Select acknowledge to clear the alarm.
• Install a fully charged rechargeable battery pack or 4 new
AA batteries. In order to start delivery, good batteries must
always be installed, even when an external source of power is
connected.
Cannot start pump with a
depleted battery.
Medium
The battery power is too low to operate the pump. In order
to start delivery, good batteries must always be installed
even when an external source of power is connected. Select
acknowledge to clear the alarm. Install 4 new AA batteries or a
fully charged rechargeable battery pack. If appropriate, start the
pump.
Cannot start pump with a
reservoir volume of zero.
Medium
The reservoir volume in the pump is set to zero. Select
acknowledge to clear the alarm. Edit the reservoir volume to
the correct value. If appropriate, start the pump.
See
Patient Specific Parameters
on page 29 for more information
about the reservoir volume.
Cannot start pump with
rechargeable battery reached
end of use.
Medium
The rechargeable battery pack is at the end of its life. It has been
discharged and recharged so many times that it is no longer
able to hold a good charge. Remove the battery from service.
In order to start delivery, good batteries must always be
installed even when an external source of power is connected.
Select acknowledge to clear the alarm. Install 4 new AA
batteries or a new rechargeable battery pack. If appropriate,
start the pump.
Cannot start pump without a
battery.
Medium
The pump does not have any batteries installed. In order
to start delivery, good batteries must always be installed
even when an external source of power is connected. Select
acknowledge to clear the alarm. Install 4 new AA batteries or a
rechargeable battery pack. If appropriate, start the pump.
Cannot start the pump with air
in-line. Prime tubing.
Medium
The air detector has detected air in the fluid path directly
under the air detector; the fluid path may contain air bubbles.
Select acknowledge to clear the alarm, then:
• If the fluid path contains air bubbles, close the clamps and
disconnect the fluid path from the patient. Then follow the
instructions for removing air by priming (see page 24 for more
information).
• If appropriate, start the pump.