
19
Pump Setup
5. Insert the pump key
into the cassette/
keypad lock and turn
clockwise into the
locked position.
NOTE
: The cassette
must be locked
in
order to start the
pump. “Cassette Locked” will
appear briefly in the status bar.
WARNING: Attach the cassette properly. A
detached or improperly attached cassette
could result in unregulated gravity infusion of
medication from the fluid container or a reflux
of blood, which could result in death or injury
to the patient.
If you are using a CADD
®
administration set or
CADD™ medication cassette reservoir that does
not have the flow stop feature: you must use a
CADD
®
extension set with anti-siphon valve
or a CADD
®
administration set with either an
integral or add-on anti-siphon valve to protect
against unregulated gravity infusion that can
result from an improperly attached cassette.
Unregulated gravity infusion can result in
death or serious injury.
Remove a Cassette
Make sure the pump is stopped before removing the
cassette.
WARNING: Per general rules of safe practice,
always clamp tubing before removing the
cassette from the pump. Removing the cassette
without closing the clamp could potentially
cause unregulated gravity infusion, which
could result in patient injury or death.
To remove the cassette:
1. Close the tubing clamp.
2. If locked, insert the
pump key and turn the
cassette/keypad lock
counter-clockwise into
the unlocked position.
“Cassette Unlocked” will briefly appear in the
status bar.
3. Push down on
the cassette latch
until the cassette
detaches.
Start the Pump
Starting the pump starts delivery. When the
pump is running, “Running” will appear with
green highlighting on the status bar and the green
indicator light will flash. If the pump will not start,
a message will appear on the display. Refer to the
Alarms and Messages table on page 36.
NOTE
: Before starting the pump, ensure correct
protocol and patient specific parameters are
displayed (see
Patient Specific Parameters
page 29).
Also be sure the tubing is primed and the pump is
connected to the patient, according to your facility’s
standards of practice.
To start the pump:
1. Press stop/start
9
.
NOTE
: If the patient
specific parameters have
not been reviewed and
the values have not been
accepted, the pump will
require you to do so
before the pump will run.
2. When “Start Pump?”
appears, select
Yes
.
Press here to select
Содержание CADD-Solis
Страница 2: ......
Страница 56: ...54 References and Troubleshooting ...
Страница 60: ...58 ...
Страница 62: ......
Страница 63: ......