Failure to properly follow warnings, cautions, and instructions
could result in death, serious injury, or damage to the pump.
Warnings
•
To avoid explosion hazard, do not use the pump in the presence of
flammable anesthetics or explosive gases.
•
If the pump is used to deliver life-sustaining medication, an
additional pump must be available. Pump failure suspends
medication delivery.
•
Use of a syringe with the CADD
®
administration set may result
in UNDERDELIVERY of medication. Syringe function can
be adversely affected by variations in plunger dimension and
lubricity, which can result in greater force required to move the
syringe plunger. A syringe plunger loses lubrication as it ages and
as a result, the amount of underdelivery increases, which could
on occasion, be significant. Therefore, the type of medication
and delivery accuracy required must be considered when using a
syringe with the CADD
®
-Solis VIP pump.
You must regularly compare the volume remaining in the
syringe to the displayed values on the pump (for example,
Reservoir Vol. and Total Given) in order to determine whether
underdelivery of medication is occurring. If necessary, contact
your health care provider.
•
Follow the instructions for use provided with the CADD
™
medication cassette reservoir, CADD
®
extension set, or CADD
®
administration set, paying particular attention to all warnings
and cautions associated with their use.
•
Carefully route tubing, cords, and cables to reduce the possibility
of patient entanglement or strangulation. Failure to observe this
warning could result in patient injury or death.