7
compartment may damage the battery contacts and could result
in loss of power and nondelivery of drug.
•
If the pump is dropped or hit, the battery door may become
broken or damaged. Do not use the pump if the battery door is
damaged because the batteries will not be properly secured. This
may result in loss of power and nondelivery of drug.
•
Do not use other AC adapters with the pump. AC adapters from
other companies can damage the rechargeable battery pack and
the pump, and could result in loss of power and nondelivery of
drug.
•
Always clamp the tubing before removing the cassette from the
pump. Removing the cassette without closing the clamp could
potentially cause unregulated gravity infusion.
•
Attach the cassette properly. A detached or improperly attached
cassette could result in unregulated gravity infusion of
medication from the reservoir, or a reflux of blood.
•
If you are using a CADD
®
administration set or CADD
™
medication cassette reservoir that does not have the Flow Stop
feature, you must use a CADD
®
extension set with anti-siphon
valve or a CADD
®
administration set with either an integral or
add-on anti-siphon valve to protect against unregulated gravity
infusion that can result from an improperly attached cassette.
•
Do not prime the fluid path with the tubing connected to your
catheter as this could result in overdelivery of medication or air
embolism.
•
To prevent air embolism, ensure that the entire fluid path is free
of all air bubbles before connecting to your catheter.
•
If the pump is dropped or hit, inspect it for damage. Do not use a
pump that is damaged or not functioning properly. If necessary,
contact your health care provider.