Turning the Pump On and Off
To turn on the pump,
press and hold the
power switch .
When the pump powers on, it performs
various self-tests. Watch and listen for
the following:
•
The green and amber lights flash.
•
The display shows a blue and
amber rotating arc, followed by a
CADD
®
-Solis Ambulatory Infusion
System display. If you see any stripes,
or black or white pixels, or if your display appears faulty, contact
your health care provider.
•
When power up completes, six beeps sound. Contact your health
care provider if you do not hear the beeps, because there may be a
problem with the audible alarms.
•
If any issues are detected during power on, alarms will sound
and/or be displayed. If the alarms continue after you follow any
on-screen help, contact your health care provider.
To turn off the pump,
press the power switch . When the
“Power
down?”
message appears, press
Yes.
The pump then powers off.
When the pump is powered off:
•
The display is blank.
•
Keypad presses are not detected.
•
Medication is not infused.
•
Alarms are not sensed, and audio and visual alarms are not
displayed.
•
If the AC adapter is plugged in, the blue AC power light remains
on. No other lights are lit.
•
If the pump is connected to AC power and a rechargeable battery
pack is installed, the battery pack continues to charge.