Using the Rechargeable Battery Pack or AC Adapter
You may have a rechargeable battery pack and an AC adapter to
power your pump. The AC adapter can be used to power the pump
and/or to recharge the rechargeable battery pack. When using
the AC adapter, the pump must also have 4 AA batteries or the
rechargeable battery pack installed as a backup.
WARNING: Do not use other AC adapters with the pump. AC
adapters from other companies can damage the rechargeable
battery pack and the pump. This may result in loss of power
and nondelivery of drug, and depending on the type of drug
being administered, could result in death or serious injury.
Note:
For complete instructions, warnings and cautions, read
the
Instructions for Use
that come with the rechargeable
battery pack or AC adapter.
To connect the AC Adapter:
1. Plug the AC power cord (the cord that
plugs into your wall outlet) into the
AC power input connector on the
AC Adapter.
2. Plug the output power cord from the
AC Adapter into the pump’s AC
power jack (labeled
“7V”
). Push the
connector firmly into the jack until
it stops. When the AC adapter is
properly connected, the blue light
next to the connector on the pump
lights. If the pump is powered on,
“AC Adapter Connected”
appears briefly on the pump display
and the battery status shows an AC power connection.
CAUTION: Failure to push the AC Adapter connector all the way
into the power jack may result in an intermittent connection
and the connector may dislodge, causing a loss of power to
the pump.
Output power cord
AC power input
AC power jack
and blue light