ALARMS MESSAGES 

There are two different situations, depending on how serious the fault is: 

 

warning messages

; there will be a message allowing the user to ignore the warning and proceed with the 

selected cycle. 

 

true alarm signals

. The user must press the “

RESET

” button and comply with the relative procedure in order 

to overcome the situation. 

By  and  large,  an  alarm  indicates  that  the  machine  is  operating  in  a  faulty  way  and  needs  to  be  repaired  by  a 
technician.  Sometimes,  however,  the  alarm  may  be  caused  by  a  temporary  situation.  Before  calling  the  authorized 
Technical Assistance service, you are therefore advised to make a RESET and repeat the cycle a second time. Call the 
authorized Technician if the alarm persists.  
When  an  alarm  message  is  noticed:  Refer  to  the  table  below  to  find  out  what  the  code  means  and  the  relevant 
countermeasures. 

Apply the countermeasures recommended for the specific circumstances.

 

 

 

DEFAULT PROCEDURE

 for all the alarms: 

1. Perform the 

RESET 

cycle. 

2. If the alarm message disappears after the 

RESET

 cycle: normal use of the device can be 

restarted. 
3. If the alarm message persists after the 

RESET

 cycle: switch the device off and back on using the 

On/Off

” main switch (Wait at least 10 seconds after switching off before switching back on 

again). 
4. If the alarm occurs when the device is restarted: retry the 

RESET

 

If a cycle has been stopped during a high temperature phase, remember that the temperature must drop 
below 45°C before the door can be opened. 

 

If the alarm persists even after the RESET procedure: 

1. Turn off the water supply taps. 
2. Disconnect the appliance from the electricity supply. 
3. Check that all the device’s connections (electricity and water) are correct and there have been no changes 
from the initial installation conditions. 

4. Contact the After-Sales Service. 

 

In case of water leakage:

 immediately Turn off the water supply taps and contact the After-Sales service.

 

 

Maintenance and any repairs must be carried out by authorized technical staff.  

For your own safety and for the safety of the device, the user activities are limited to control activities, 
external to the device, and to the activation of the RESET cycle, if requested. 

 
 

ALARM ID 

DISPLAY MESSAGE 

MEANING AND SUGGESTED USER ACTION 

ID1 

“Water heating failed” 

Water not heated within time allowed. 
 

.Check the condition of the safety thermostat (push the first red 
plunger on the upper wall of the detergent cabinet to reset the 
thermostat).  

ID2 

“Temp. Probe TL1-TC” 

The temperature difference between the two probes, “TL1” and 
“TC”, is more than 2°C  
 
.

follow the 

DEFAULT PROCEDURE

 described above.  

ID4 

“Overtemperature TL1” 

Probe “TL1” (chamber temperature) reads a higher temperature 
than the target one. 
 

.follow the 

DEFAULT PROCEDURE

 described above.  

Содержание WD3060

Страница 1: ...WD3060 WD5090 WD3060 WD5090 WASHER DISINFECTOR USER MANUAL...

Страница 2: ...CTING CYCLE 14 7 4 HALTING A PROGRAM IN PROGRESS 15 7 5 PERFORMING A STAND ALONE DRYING CYCLE 15 7 6 PRINTING OUT THE LAST CYCLE PERFORMED 16 7 7 POSTPONING THE CYCLE START 16 7 8 SECURITY POLICY USER...

Страница 3: ...IS TO BE OUT OF USE 32 13 4 REUSING THE DEVICE AFTER A PERIOD OUT OF USE 32 13 5 TROUBLESHOOTING 33 14 ROUTINE INSPECTIONS AND MAINTENANCE 34 14 1 DAILY 34 14 2 WEEKLY 34 14 3 EVERY SIX MONTHS 34 14...

Страница 4: ...EC 61010 1 2001 The manufacturer declines all liability for uses other than those described in this manual Consumables detergents air filters thermal printer paper etc are not covered by the warranty...

Страница 5: ...ts unless the electricity supply is disconnected Transport and unpacking precautions Biohazard Flammability hazard At the end of its working life the product must be consigned to a disposal plant for...

Страница 6: ...by specifically trained staff The authorized installation engineer will train the staff assigned to use the appliance at the time of installation The manufacturer declines all liability in the event...

Страница 7: ...redetermined limits when used in accordance with its operational procedures performance qualification PQ process of obtaining and documenting evidence that the equipment as installed and operated in a...

Страница 8: ...ump Drying system HEPA H13 absolute filter air Steam condenser P1 Detergent dispensing monitoring P2 Neutralising agent dispensing monitoring P3 pump dispensing monitoring P4 pump dispensing monitorin...

Страница 9: ...is not designed for users to climb into it The user must never climb into the chamber this might put his safety at risk ref 7 102 IEC61010 2 040 2005 DEMI WATER ABSENCE If demineralised water is not a...

Страница 10: ...s or performing any maintenance Perform a complete disinfection cycle or if you cannot perform a complete Thermo disinfection cycle handle tools with caution using protection devices required for the...

Страница 11: ...The door of the side cabinet can be open simply pulling its handle 6 2 CONTROLS The keyboard is divided into two separate sections The left hand part of the display with keys for starting and stoppin...

Страница 12: ...s no Symbols similar to the keys appear on the bottom line of the display as their functions are activated Some keys have special functions depending on the active screen such as the STOP key if used...

Страница 13: ...be opened manually by releasing the lock using a screwdriver with stem 3mm taking care not to damage the device 1 This procedure is only possible with the appliance in stand by status with no cycle ru...

Страница 14: ...1 10 00 Drying On 05 02 2013 08 30 15 Fig 5 Screen of the program selected Once the program has started the display shows the following information the name of the program being run e g PLAST 70 C the...

Страница 15: ...y program except for prewash programs by default To start a stand alone drying phase Press LEFT from the menu of the selected program The drying phase activation screen appears press ENTER to confirm...

Страница 16: ...Set the delay time using the R and L selector buttons to locate the cursor on the hours and minutes and the INC and DEC buttons to increase or decrease the value PR 3 PLASTIC 70 C As 000003 Phases N...

Страница 17: ...CTIVATION DETERGENT DISPENSERS FLOWMETER CONDUCTIVITY SENSOR DWP ACTIVATION TEMPERATURE ADJUSTMENT MAINTENANCE Manufacturer level concerns the manufacturer of the appliance Default Passwords USER PASS...

Страница 18: ...n Carry out a prewash cycle to prevent this if necessary 8 2 USING DETERGENTS One fundamental factor for a good washing process is the type of detergents used The manufacturer guarantees excellent was...

Страница 19: ...t safety data sheets Once a product has run out replace the empty can with a full can of the same product If the product left in the old can is poured into the new one take care not to overfill the ne...

Страница 20: ...lf Turn the adjuster lever fully towards the left normal position to obtain the maximum pressure in the top spraying arm or in the injection system Turn the lever towards the right towards the bottom...

Страница 21: ...time intervals between this preliminary treatment and the instrument washer treatment as the risk of corrosion exists for both methods The instruments must be placed appropriately in instrument holder...

Страница 22: ...a program it is advisable to run program N 1 chamber prewash empty chamber at the beginning of a work day as a preventive safety measure A washing program generally comprises phases of different types...

Страница 23: ...ept constant at close to the target value set Programs that include thermal disinfection have therefore been designed to offer the following A0 values Temperature C and time min A0 90 C 1 600 90 C 5 3...

Страница 24: ...aining pump P1 P2 P5 detergents peristaltic pumps PAP high pressure switch washing pump ML monitoring PL1 working level pressure switch PS safety level pressure switch PSC steam condenser draining pum...

Страница 25: ...tching back on again 4 If the alarm occurs when the device is restarted retry the RESET If a cycle has been stopped during a high temperature phase remember that the temperature must drop below 45 C b...

Страница 26: ...tive presence absence of warm water ID13 Lack of demi water The same as ID11 for the demi water Check the correct setting of the parameters of the device effective presence absence of demi water ID17...

Страница 27: ...ter hardness hard water contribute to steam condenser nozzles obstruction ID34 Condenser Drainage failed Steam condenser water drainage failure Check that the drain connections are correct 1 Height an...

Страница 28: ...upper wall of the detergent cabinet to reset the thermostat ID60 Drying 1 Safety Switch on Drying heating system failure Check the condition of the safety thermostat push the second red plunger on th...

Страница 29: ...r open with a cycle in progress always use the buttons provided on the appliance to stop a cycle and open the door 3 Check there is nothing between the door and the appliance s chamber preventing the...

Страница 30: ...device must be cleaned at regular intervals once a month is recommended with a soft cloth soaked in water or a normal detergent for steel surfaces The door seals must be cleaned with a damp sponge Aft...

Страница 31: ...biologically contaminated material and must be handled with due care using protective equipment e g gloves protective glasses lab coat The potentially contaminated material in the filters and any oth...

Страница 32: ...year and to visually check to see how much the second filter has darkened at this time The appliance will warn when the absolute filter needs replacing A clogged filter contributes to the degradation...

Страница 33: ...fail to dry or are dull make sure that Absolute filter condition operating hours can be viewed as an appliance parameter the default operating hours are set for normal conditions of use the drying fil...

Страница 34: ...s not include the replacement of components which deteriorate due to normal wear and tear The operations which the authorised staff will carry out are 1 Inspection and if necessary replacement of worn...

Страница 35: ...____________________________________ Accessories present on the machine place a check next to the accessory present and write the quantity provided CONFIGURATION TABLE PERISTALTIC PUMPS N 1 2 3 4 5 DE...

Страница 36: ...with guidance about keeping your appliance functioning correctly and put you in touch with your nearest authorized Service Centre INTERNATIONAL CUSTOMERS PLEASE CONTACT YOUR LOCAL DISTRIBUTOR Smeg S p...

Отзывы: