Revision n.1
User Manual B-One
Edition: 04/2009
Page 19 of 52
Completed cycle sequence number (ex. “CYCLE N. 00029” image);
current cycle name (ex. “CLASS B 134 °C”);
chamber temperature (ex. “56.1 °C”)
chamber pressure (ex. “1.01 bar”);
instantaneous progress graphic with indication of time elapsed;
“lock” icon indicating the door status, locked:
preheating icon (if the icon image is not present, preheating has been completed);
“STOP” icon, for forced cycle interruption
5.4
AVAILABLE CYCLES – STERILIZATION AND TESTS
Available sterilization cycles:
•
Class B 134°C
•
Class B 121°C
•
Fast 134°C
•
Prion 134°C
2 test cycles:
•
Vacuum test (auto vacuum seal test)
•
B&D / helix test (class B auto-test)
The following table summarises sterilization cycle features.
Cycle
Number
of
vacuums
Sterilization
temperature
Sterilization
time
Drying time
Description
121°C
16:00
CLASS B
4
134°C
4:00
9:00
Cycle for
sterilization of
turbines,
handpieces,
hollow bodies,
porous, double
bagged
FAST
3
134°C
4:00
2:00
Cycle for
sterilization of
solid,
unbagged
instruments,
not suitable for
hollow bodies,
type N
PRION
4
134°C
22:00
9:00
Prion cycle