4. If the error persists: switch off the machine, wait for components to cool
down for at least 30 minutes, switch machine back on. Component overheating
problems, generally caused by improper equipment use/positioning, can be
resolved in this way. We recommend checking autoclave positioning and re-
reading manual parts related to proper use.
5. Contact technical assistance if alarms persist.
FIRST
GROUP
SECOND
GROUP
ALARM DESCRIPTION
E1
1
Sudden shutdown / power loss experienced during
wake time programming.
E1
2
Door locking device alarm.
E1
4
Electronic time scanning check of time has failed.
E1
8
Heating resistance problem: electronic feedback not
compliant with command given.
E1
16
Vacuum test cycle failed.
E1
32
Timeout during machine operations - said alarm is
better specified in the THIRD GROUP of characters
(see following table).
E1
64
Steam generator water level too low.
E1
128
Pressure probe generating a value outside of allowed
limits.
E2
1
Temperature probe generating a value outside of
allowed limits.
E2
2
Electronic problem during writing of parameters of
memory.
E5
1
Knob not detected by the programme unit.
E5
2
Interface board – microcontroller communication fault.
E5
3
Interface board – flash memory not programmed
E5
4
Interface board – communication error between the
interface and the cpu board.
E5
6
SD card not connected.
E5
7
SD card format not known.
E5
8
SD card full.
E5
10
Cpu board – jumper off
E5
12
Demi switch position not correct