RECOMMENDED SUPPLY WATER FEATURES
RESIDUES (EXCEPT
IRON, CADMIUM, LEAD)
CHLORIDES
≤
2 mg / l
PHOSPHATES
≤
0.5 mg / l
CONDUCTIVITY AT 20°C
≤
15
µS/cm
S/cm
pH LEVEL
From 5 to 7.5
APPEARANCE
Colourless, transparent,
without sediments
HARDNESS
≤
0.02 mmol / l
5.8.4 MANUAL DRAINING OF LOADING TANK WATER
The loading tank can be drained manually for transport and maintenance needs and any
other time that input water quality is indicated as unsuitable for sterilization.
Manual draining procedure:
1. Set the free end of the supplied tube in a large container for the water to be drained
into, set underneath the autoclave.
2. Connect the end with the tube connection on the corresponding quick couple (gray)
on the front part of the equipment (the second quick couple from the right). Position
the tube in such a way to prevent any blockages. The quick connect is indicated
with an icon similar to the "unsuitable quality" icon for supply water.
3. Let the water flow entirely into the container
5.8.5 WATER SUPPLY TANK DRAINING
During sterilization, used water is collected in a suitable tank ("discharge tank").
If the level exceeds the maximum allowed level, the machine will warn the operator with a
message (“DISCHARGE TANK FULL").
fig. 21 – “DISCHARGE TANK FULL” warning screen