Revision n.1
User Manual B-One
Edition: 04/2009
Page 26 of 52
The machine will not allow any new sterilization cycles to be carried out until the
tank is drained.
To drain the tank:
1. Set the free end of the supplied tube in a large container for the water to be drained
into, set underneath the autoclave.
2. Connect the end with the tube connection on the corresponding gray quick couple
on the front part of the equipment (the second quick couple from the right). The
quick connect is indicated with an icon similar to the "discharge tank full" icon.
3. Press the knob
4. Let the water flow entirely into the container.
If necessary, tank draining can be interrupted by pressing the knob a second time.
5.8.6 AUTOMATIC DRAINING
When a free discharge connection is set near the machine, authorised Smeg technical
personnel can configure discharge in such a way that it is carried out automatically,
without the user needing to manually manage operations.
The hydraulic system connection for used water discharge must be set at a height which is
lower than the autoclave support level and must not allow for any water reflux to the
machine.
The maximum allowed distance from the hydraulic connection to the connection set on the
machine is 1.5m.
5.8.7 WATER PURIFIER WATER SUPPLY
The Smeg DEMI-One water purifier is an optional accessory for the autoclave.
DEMI-One is designed to produce quality demineralised water.
DEMI-One connection to the autoclave must be carried out by authorised Smeg
technical personnel.
DEMI-One can operate in two modes: automatic and semi-automatic.
Automatic mode: autoclave tank top up is managed completely automatically by the
equipment itself.
Semi-automatic mode: the user is requested confirmation for tank top up via the display.
Simply press the knob for top up.