[F22] Setting of analogue output
This function is available when the model includes the analogue output.
The flow value corresponding to an analogue output voltage = 5 V or analogue output current = 20 mA can
be selected.
: When using a model without analogue outputs, this setting is not available and [---] will be displayed.
<Operation>
Press the
or
button in function selection mode to display [F22] on the main display.
[FRE] and [the current set value] are displayed in turn on the sub display.
Press the
button.
Setting of analogue output
[FrE] is displayed on the main display.
[The current set value] flashes on the sub display.
Press the
or
button to select.
Select [oFF]. Select [oN].
Press the
button to set. Press the
button to set.
Set value
[F_H] is displayed on the main display.
[The current set value] flashes on the sub display.
Use the
and
buttons to enter the flow
value that will generate
full span analogue output
(5 V or 20 mA).
The entered flow value
can be in the range: 10%
of the maximum rated
flow, to the max.
displayable range.
Press the
button to set.
[F22] Setting of analogue output completed.
Return to function selection mode.
: The above example is for the 500 L/min type.