P
Peak value display (mode)
Shows the maximum pressure from when the power was supplied to the
current time.
Port size
The diameter of the connecting part of the switch for connecting with the
object to be measured.
Power saving mode
Operating mode in which the digital display turns off and power consumption
is reduced.
Pressure setting
The set pressure value that determines the point at which the pressure switch
turns ON and OFF.
Proof pressure
Pressure limit that if exceeded will result in mechanical and/or electrical
damage to the product.
R
R.D.
Current read value
For example, when the display value is 1.000[MPa],
±
5%R.D. is
±
5% of
1.000[MPa], which becomes
±
0.05[MPa]. When the display value is
0.800[MPa],
±
5%R.D. is
±
5% of 0.800[MPa], which becomes
±
0.04[MPa].
Rated pressure range
The pressure range within which the product will meet all published
specifications.
Values outside of this range can be set as long as they are within the set
pressure range, but the specifications cannot be guaranteed.
Repeatability
Variation in repeated measurement of pressure display or ON-OFF output
point when the pressure changes at 25 centigrade.
Residual voltage
The difference between the ideal ON voltage and the actual voltage when the
switch output is on. Varies with load current. Ideally should be 0 V.
Resolution
Refer to "Display resolution".
Reversed output
One of the switch output types. In hysteresis mode the switch output is turned
ON when pressure less than or equal to the switch output set value is
detected. In window comparator mode, the switch output is turned ON when
pressure is outside the switch output set values (n1L to n1H) is detected.
(Refer to the "List of output modes" on page
.)
Ripple
A type of chattering.
S
Set pressure range
The pressure range that can be set for switch output.
Switch output
Sometimes referred to as "ON-OFF output".
U
Units selection function
A function to change the units in which the measured pressure value is
displayed. The display units can only be changed if the product is equipped
this function. It is not possible to purchase the product with this function if the
product is used in Japan.
The product for Japan is displayed in SI only.
Содержание ISE20CH-L
Страница 96: ...95 No PS OMX0002 Dimensions 02 type...
Страница 97: ...96 No PS OMX0002 F02 type C01 type A2 type B2 type...
Страница 98: ...97 No PS OMX0002 02L type...
Страница 99: ...98 No PS OMX0002 F02L type C01L type A2L type B2L type...
Страница 100: ...99 No PS OMX0002 Bracket mounting dimensions Bracket A...
Страница 101: ...100 No PS OMX0002 Bracket C...
Страница 103: ...102 No PS OMX0002 Mounting dimensions of panel mount adapter bottom piping type...
Страница 104: ...103 No PS OMX0002 Mounting dimension of panel mount adapter bottom piping type Front protective cover...
Страница 106: ...105 No PS OMX0002 Vertical...