Chapter 8 Control, Inspection and Cleaning
HRS-R Series
8.2
Inspection and Cleaning
8-3
8.2.2
Monthly check
Table 8-3 Contents of Monthly check
Item
Content of check
Ventilating condition
Clean the
ventilation grille.
Make sure the ventilation grille is not clogged with
dust, etc.
Automatic water supply
(Option J: Automatic
water supply)
Check the supply
water.
Make sure the suppy water is clean and contains no
foreign matter.
Cleaning of air vent
Removal of the dustproof filter
1.
The dustproof filter is installed at the lower part of the front face of the thermo-chiller.
2.
Slide the dustproof filter upwards to remove.
Care should be taken not to deform or scratch the air-cooled condenser.
Fig. 8-1 Removal of the dustproof filter
1
2
If the fins of the air-condenser become clogged with dust or debris, heat
radiation performance reduces. This results in the reduction of cooling
performance, and may stop the operation because the safety device is trigger.
Shut off the power supply of the product when performing cleaning,
maintenance or inspection. It might cause electric shock, injury or burn, etc.
Replace all panels removed for inspection or cleaning. It might cause injury or
electric shock if it is operated with the panel removed or opened.
Use a long bristled brush or air gun to clean the condenser to prevent the fins
from being deformed or damaged.
Содержание HRS018-A
Страница 16: ...HRX OM W058 Chapter 1 Safety Instructions 1 8 Material Safety Data Sheet MSDS HRS R Series 1 10...
Страница 22: ...HRX OM W058 Chapter 2 Name and Function of Parts 2 4 Operation display panel HRS R Series 2 6...
Страница 50: ...HRX OM W058 Chapter 3 Transport and Setting Up 3 7 Wiring of external switch HRS R Series 3 28...
Страница 130: ...HRX OM W058 Chapter 6 Option Optional Accessories 6 5 Option V Stainless steel enclosure HRS R Series 6 4...
Страница 146: ...HRX OM W058 Chapter 8 Control Inspection and Cleaning 8 4 Stop for a Long Time HRS R Series 8 10...
Страница 161: ......