RotaScope Classic Cabinet
Manual Version
1.04
Date
23 November 2021
53
3.3.2
Scope Connector Hose Kits
–
Cleaning IFU
Scope Connector Hose Kits
and ESC connectors (Endoscope Connectors) are to be
disinfected every 7 days or upon disconnection from a valid stored endoscope if time
since last disinfection exceeds 7 days. Hoses and ESC connectors should be inspected
for any damage or degradation of the items that would diminish the
items’
ability to
perform at its full capacity. Before attaching the hose kit to the scopes or the cabinet
we recommend manually disinfecting the hoses and lines. Alternatively, the hose kits
may be autoclaved as per your local guidelines.
RECOMMENDED STEPS for Manual Cleaning Scope Connector Hose Kits:
(IFU) Instructions For Use
1.
Prepare appropriate amount of cleaning solution (biofilm remover or
enzymatic detergent solution) in a sink, as per the chemical
manufacturer’s instructions.
2.
Remove hose/s to be cleaned from cabinet and place in sink.
3.
Make sure the hoses remain fully immersed throughout the cleaning
process.
4.
Use a small soft brush to clean the connectors at each end of the hose.
5.
Wipe exterior surfaces of hoses with a nonabrasive and lint free cloth.
6.
Check that all visible debris have been removed.
7.
Optionally, attach hoses to a
Washer Rack Hose Connection Kit
.
8.
Use a syringe to push fresh solution through hoses and then flush out.
Repeat a minimum of two more times, do not flush out on last action.
9.
Achieve contact time with hoses submerged for the appropriate
timeframe outlined in the chosen cleaning solution IFU.
10.
Purge cleaning solution from channels while fully immersed.
11.
Repeat steps 7-10 until all hoses have been flushed and purged.
12.
Discard cleaning solution.
13.
Use a syringe to push fresh water through hoses to remove all cleaning
solution.
14.
Thoroughly wash exterior of all hoses with fresh water to remove all traces
of cleaning solution.
15.
Remove hoses from sink and hold in a vertical position to allow water to
drain out. Use a syringe or medical air supplied air gun to purge any
remaining water from the inside of the hose.
a.
Dry in a drying cabinet where possible.
b.
Dry with medical grade air where possible.
16.
Remove hoses from the
Washer Rack Hose Connection Kit.
Hang hoses back inside endoscope cabinet connected to forced air for
additional drying.
This page may be printed and displayed at your wash area
Содержание RotaScope Classic Series
Страница 96: ...RotaScope Classic Cabinet Manual Version 1 04 Date 23 November 2021 96 CE Declaration of Conformity 2021 P1 ...
Страница 97: ...RotaScope Classic Cabinet Manual Version 1 04 Date 23 November 2021 97 CE Declaration of Conformity 2021 P2 ...
Страница 99: ...RotaScope Classic Cabinet Manual Version 1 04 Date 23 November 2021 99 Workflow Overview ...
Страница 100: ...RotaScope Classic Cabinet Manual Version 1 04 Date 23 November 2021 100 Barcode Setup Overview ...