HI302 - User’s Manual
4.3
Command-Specific Response Codes
Code
Class
Description
0
Success
No Command-Specific Errors
1-5
Undefined
6
Error
Device-Specific Command Error
7
Undefined
8
Warning
Update Failure
9-15
Undefined
16
Error
Access Restricted
17-127
Undefined
Command 18 - This command writes the Tag, the Descriptor and the Date into the
device.
Request Data Bytes
Byte
Format
Description
0-5
Packed
Tag
6-17
Packed
Descriptor Used By The Master For Record Keeping
18-20
Unsigned-24
A Date Code Used By the Master For Record Keeping
(e.g. Last or Next Calibration Date).
Response Data Bytes
Byte
Format
Description
0-5
Packed
Tag
6-17
Packed
Descriptor
18-20
Date
Date Code
Note:
The values returned in the response data bytes reflect the value actually used by
the Field Device.
Command-Specific Response Code
Code
Class
Description
0
Success
No Command-Specific Error
1-4
Undefined
5
Error
Too few data bytes received
6
Error
Device-Specific Command Error
7
Error
In write protect mode
8
Undefined
9
Error
Invalid data code detected
8-15
Undefined
16
Error
Access restricted
17-31
Undefined
32
Error
Busy
33-127
Undefined
Note:
Most field devices only store the data. As a result, some field devices may not
detect an invalid date code.
Содержание HI302
Страница 1: ...AGO 02 FOUNDATION H I 3 2 F M E...
Страница 4: ...HI302 User s Manual IV...
Страница 10: ...General Vision 1 4...
Страница 20: ...Installation 2 10...
Страница 46: ...Operation 5 12...
Страница 57: ...HI302 User s Manual 7 5...
Страница 59: ...HI302 User s Manual 7 7...
Страница 60: ...An Example of HI302 Usage 7 8...
Страница 64: ...Troubleshooting 8 4...