HI302 – User’s Manual
C.8
HVT ALLOCATION MAP FOR SMAR LD301 V6.01
Index
Parameter Name
HART Variable Name
HART Command
Description
14.12
U8B_ARRAY_4[12]
WRITE_SENSOR
153
w
Write Sensor
16.1
FLOAT_ARRAY_1[1]
PV_CURR_LEVEL_DAC_ZERO
45
w
Set value of the Trim Primary Variable Current Dac Zero
16.2
FLOAT_ARRAY_1[2]
UPPER_SENSOR
128
r
Upper Sensor Value
16.3
FLOAT_ARRAY_1[3]
LOWER_SENSOR
128
r
Lower Sensor Value
16.4
FLOAT_ARRAY_1[4]
UPPER_TRIM_POINT
130
w
Write Upper Sensor Trim Point
16.5
FLOAT_ARRAY_1[5]
LOWER_TRIM_POINT
131
w
Write Lower Sensor Trim Point
16.6
FLOAT_ARRAY_1[6]
MEASURED_POINT_1
160,162
w/r
Measured characterization Trim Curve Point 1
16.7
FLOAT_ARRAY_1[7]
MEASURED_POINT_2
160,162
w/r
Measured characterization Trim Curve Point 2
16.8
FLOAT_ARRAY_1[8]
MEASURED_POINT_3
160,162
w/r
Measured characterization Trim Curve Point 3
16.9
FLOAT_ARRAY_1[9]
MEASURED_POINT_4
160,162
w/r
Measured characterization Trim Curve Point 4
16.10
FLOAT_ARRAY_1[10]
MEASURED_POINT_5
160,162
w/r
Measured characterization Trim Curve Point 5
16.11
FLOAT_ARRAY_1[11]
PROCESS_VARIABLE
140
r
Process Variable (PV) Value
16.12
FLOAT_ARRAY_1[12]
SET_POINT
140,146
w/r
Set Point (SP) Value
16.13
FLOAT_ARRAY_1[13]
MANIPULATED_VARIABLE
140,147,193
w/r
Manipulated Variable (MV) Value
16.14
FLOAT_ARRAY_1[14]
ERROR
140,193
r
Error Value
16.15
FLOAT_ARRAY_1[15]
PROPORTIONAL_FACTOR(Kp)
142,143
w/r
Proportional Factor (Kp) Value
16.16
FLOAT_ARRAY_1[16]
INTEGRAL_TIME(Tr)
142,144
w/r
Integral Time (Tr) Value
16.17
FLOAT_ARRAY_1[17]
DERIVATIVE_TIME(Td)
142,145
w/r
Derivative Time (Td) Value
16.18
FLOAT_ARRAY_1[18]
NON_LINEAR_FACTOR(Knl)
142
r
Non Linear Factor (Knl) Value
16.19
FLOAT_ARRAY_1[19]
DERIVATIVE_FACTOR(DG)
142
r
Derivative Factor (DG) Value
16.20
FLOAT_ARRAY_1[20]
MV_HIGH_LIMIT
148,149
w/r
Manipulated Variable High Limit
17.1
FLOAT_ARRAY_2[1]
MV_LOW_LIMIT
148,149
w/r
Manipulated Variable Low Limit
17.2
FLOAT_ARRAY_2[2]
MV_RATE_CHANGE
148,149
w/r
Manipulated Variable Rate of Change
17.3
FLOAT_ARRAY_2[3]
POWER_UP_SETPOINT
150,151,194
w/r
Power Up Set Point
17.4
FLOAT_ARRAY_2[4]
POWER_UP_MANIPULATED_VARIABLE 150,151
w/r
Power Up Manipulated Variable
17.5
FLOAT_ARRAY_2[5]
CUTOFF_POINT
156,157
w/r
Square Root Cutoff
17.8
FLOAT_ARRAY_2[8]
USER_UNIT_UPPER
178,179
w/r
User Unit Upper Value
17.9
FLOAT_ARRAY_2[9]
USER_UNIT_LOWER
178,179
w/r
User Unit Lower Value
17.10
FLOAT_ARRAY_2[10]
TOTAL
185
r
Total Value
17.11
FLOAT_ARRAY_2[11]
MAXIMUM_FLOW
186,187
w/r
Maximum Flow Value
17.12
FLOAT_ARRAY_2[12]
TOTAL_UNIT_CONV_FACTOR
186,188
w/r
Total Unit Conversion Factor Value
17.13
FLOAT_ARRAY_2[13]
PV_CURR_LEVEL_DAC_GAIN
46
w
Set value of the Trim Primary Variable Current Dac Gain
17.14
FLOAT_ARRAY_2[14]
SPECIAL_UNIT_OUT
193
r
Special Unit Output Value
17.15
FLOAT_ARRAY_2[15]
SET_POINT_VARIABLE
193
r
Set Point Variable Value
Содержание HI302
Страница 1: ...AGO 02 FOUNDATION H I 3 2 F M E...
Страница 4: ...HI302 User s Manual IV...
Страница 10: ...General Vision 1 4...
Страница 20: ...Installation 2 10...
Страница 46: ...Operation 5 12...
Страница 57: ...HI302 User s Manual 7 5...
Страница 59: ...HI302 User s Manual 7 7...
Страница 60: ...An Example of HI302 Usage 7 8...
Страница 64: ...Troubleshooting 8 4...