![SMAR FI303 Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/smar/fi303/fi303_operation-maintenance-and-instruction-manual_1294912039.webp)
Configuration
3.23
W1 Jumper Connections
If W1 is connected to ON, the local adjustment programming tree is enabled and then important
block parameters can be adjusted and communication can be pre-configured via local adjustment.
Fig. 3.19 - J1 and W1 Jumpers
Example: let's say we want to calibrate the lower and upper current value. From normal display,
enter local adjustment. The display will show:
Figure 3.20 - Step 1 - FI303
MAIN BOARD
WR
J1
OFF
ON
W1
LOC
ADJ
OFF
ON
In order to start the
local adjustment,
place the magnetic
tool in orifice
Z
and
wait until letters
MD
are displayed
.
Place the
magnetic tool in
orifice
S
and wait
during 5
seconds.