![SKC AirChek TOUCH 220 Series Скачать руководство пользователя страница 3](http://html1.mh-extra.com/html/skc/airchek-touch-220-series/airchek-touch-220-series_operating-instructions-manual_1286847003.webp)
Form 38054 Issue 1506
Page 3 of 30
Notes and Cautions
• Do not operate pump from or charge pump with charging cradle in hazardous locations.
• Power off pump before removing battery to avoid loss of time, date, and other settings.
• Use only the SKC-approved battery pack (Cat. No. P75718) and charging cradle (Cat. No. 220-800 or 220-900) for pump. Use
of an unapproved battery and/or charging cradle could damage the pump and will void any warranty.
• Tampering with the battery pack (opening, disassembling, short circuiting, crushing, or exposing the battery pack to
fi
re or
temperatures in excess of 212 F [100 C]) voids any warranty.
• User may replace external components such as the inlet
fi
lter, battery, protective screen cover, and/or belt clip. Service must
be done by SKC to maintain performance and IS rating. Warranty is void if pumping compartment is opened by user.
• Failure to follow warnings and cautions voids any warranty.
• WARNING: Substitution of components may impair intrinsic safety. ADVERTISSEMENT: La substitution de composants peut
compromettre la Sécurité Intrinsèque.
• CAUTION: The battery used in this device may present a risk of
fi
re or explosion when heated above 212 F (100 C) or incinerated.
Replace battery with SKC Unlisted Component Battery Pack model P75718 only. Use of another battery may present a risk of
fi
re or explosion.
• Model 220-5000TC: Exia – Intrinsically safe/ Sécurité Intrinsèque
• Warning: To prevent ignition of a hazardous atmosphere, batteries must only be changed [removed and replaced] in an area
known to be non-hazardous. ADVERTISSEMENT: A
fi
n de prévenir l’in
fl
ammation d’atmosphères dangereuses, ne changer les
batteries que dans des emplacements désignés non dangereux.
• CAUTION: Risk of Fire and Burns. Do Not Disassemble, heat above 212 F (100 C), or incinerate. Keep battery out of reach of
children and in original package until ready to use. Dispose of used batteries promptly according to [all state and] local recycling
or waste regulations.
Turning Pump Power On/Off (Figure 1, K)
Turn on:
Press and hold brie
fl
y the recessed power bu
t
on on the side of the pump [Figure 1, K]. The screen will light.
Turn o
ff
:
Press and hold brie
fl
y the power bu
t
on. The screen will turn o
ff
.
Note:
To conserve ba
t
ery power, a non-running pump will power o
ff
automatically after 5 minutes of inactivity. Also see Auto Dim feature/se
t
ing
in Modify Device Se
t
ings, Security.
Determining Battery Charge Status
The ba
t
ery status icon at the top right of the Home screen contains four bars that reduce in number as ba
t
ery charge is depleted. Use the
table below to interpret the ba
t
ery status icon.
Icon Displayed
Battery Charge Remaining
Four bars
Full battery charge, approximately 75 to 100%
Three bars
Approximately 50 to 75%
Two bars
Approximately 25 to 50%
One bar
Approximately 5 to 25%
No bars
Low battery fault is imminent. Pump will stop and power off eventually. Run time data will be retained in history.
A fault icon will appear on the screen once the pump is restarted.