Form 38054 Issue 1506
Page 11 of 30
3. Set up a calibration train. See Figure 7.
4. Touch up and down arrow keys to adjust calibration based on
fl
ow displayed on
calibrator. Touch checkmark to move to Manual Cal Completed screen.
5.
Touch checkmark to accept calibrated
fl
ow; return to Home screen. Touch X to cancel
fl
ow adjustment; return to Home screen.
6. Disconnect pump from representative method-speci
fi
ed sampling medium and
calibrator and proceed to
Sampling
.
Alternative Manual
Calibration Access
Home
Settings
Calibrate
Manual
Connect calibration
train
checkmark
enter
fl
ow
rate
checkmark
use arrows
to adjust
fl
ow
checkmark
checkmark to accept or X to cancel
Home screen
Figure 7.
Calibration Train (1 to 5 L/min)
Defender calibrator
Tubing
Representative
sample medium
Tubing
Pump inlet