Operation
Setting Flow Rate
1. Touch flow rate display (Set
Flow soft key) on a pump that
is
not
running. Pump will run
and display Set Flow screen.
2.
Touch up/down arrow soft keys*
to adjust flow rate
or
touch
keypad soft key to enter desired
fl
ow rate and proceed to
Manual
Calibration
.
* Note:
Using the up/down arrow soft keys to adjust
fl
ow will allow access to manual calibration from this screen. Touch
fl
ow rate display (Manual
Cal soft key – tools icon in upper right corner).
Tip:
Tapping up and down arrow soft keys increments/decrements
fl
ow rate in steps of 1. Touching and holding the soft keys speeds
increment/decrement to the next 10 and then proceeds in steps of 10.
Calibrating Flow Rate from 1 to 5 L/min
• Allow pump to equilibrate after moving it from one temperature extreme to another.
• Charge pump battery completely before calibration and sampling.
• Choose from Manual or CalChek Single calibration methods.
Manual Calibration
1. Touch
fl
ow rate display (Set Flow soft key) on a pump that
is
not
running. Pump will run and display Set Flow screen.
2. Touch up/down arrows to set desired
fl
ow rate
(see Se
t
ing
Flow Rate)
. Allow pump to run for
fi
ve minutes. Touch
fl
ow
rate display (Manual Cal soft key – tools icon in upper right
corner).
Checkmark
accepts set
fl
ow
and returns to Home screen.
Back
moves to Home screen
without accepting set
fl
ow.
Manual Cal (flow rate
display/Manual Cal soft
key)
accesses Manual Cal
screen
(see Calibrating
Flow Rate from 1 to 5 L/min,
Manual Calibration to set up
calibration train).
Select option: