Enter the total duration of the sample run, and then specify number of minutes the pump is to run and number of
minutes the pump is to pause during a single cycle. The AirChek TOUCH will calculate the number of run/pause
cycles and the estimated volume.
Note:
During calculations, the pump may make slight adjustments to the total duration
to
fi
t in the desired number of run/pause cycles.
Intermittent Sample
1. Touch
Intermi
t
ent.
2. E n t e r t o t a l d u r a t i o n o f
sample run in minutes. Touch
checkmark to accept. P1 Cycle
Sample screen displays.
4.
Enter number of minutes pump
is to pause between times
of active sampling. Touch
checkmark to accept. P1 Start
Time screen displays.
5. a.
To bypass se
t
ing a start time:
Touch No. Preset 1 screen
displays.
b.
To enter a start time:
Touch
Yes. P1 Start Time screen
displays.
Date
touch to edit start date.
Run
starts sampling. Screen will lock automatically (regardless of
security settings).
See Options/Modes During Sampling.
Back
moves to Timed Sample screen.
Select option:
3.
Enter number of minutes pump
is to actively sample during
each cycle. Touch checkmark to
accept. P1 Cycle Pause screen
displays.
5.b.iEnter a start time (hh:mm),
touch am/pm/24 hour to select
type of time display, and touch
checkmark to accept.
Edit
moves to Flow Rate screen; allows
fl
ow rate, type of sample,
total duration, cycle sample, cycle pause, and start time to be
changed.
Note:
Touch Erase soft key on each screen to erase
existing values and enter new values.
Home
moves to Home screen.