![SKC AirChek TOUCH 220 Series Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/skc/airchek-touch-220-series/airchek-touch-220-series_operating-instructions-manual_1286847002.webp)
Form 38054 Issue 1506
Page 2 of 30
Pump
DATA
Charge
status
Charging
contacts
Charging
cradle
Power
port
Single cradle
power supply
100-240 V
Wall outlet
Multi cradle
power supply
Figure 2.
Charging Train, Single and Multiple Cradles
LED Action
Charge Status
Red
steady
Charge in progress
Red
3 sec
Green
1 sec
(Pattern
repeats)
Approximately 75% charged
Green
steady
Charge completed/trickle charge
Getting Started
Charging the Battery Pack
Set up charging train (Figure 2) and completely charge ba
t
ery pack(s) before operating the pump.
1. Prepare charging cradle(s).
a.
Single cradle:
Insert connector on Single Cradle Power
Supply (Cat. No. 220-600) into power port on side of
Standard Charging Cradle (Cat. No. 220-800) or Charging
e-Cradle (Cat. No. 220-900). Insert wall cube into a 100 to
240-volt wall outlet.
b.
Up to 5 cradles:
Press together the connector on the side
of the
fi
rst cradle with the connector on the side of the
succeeding cradle. Repeat connection to chain up to 5
Standard Charging Cradles or up to 4 Standard Cradles
and 1 Charging e-Cradle. Insert connector of Multi Cradle
Power Supply (Cat. No. 220-700) into power port on side
of the last cradle in the chain. Insert wall cube into a 100
to 240-volt wall outlet.
2.
Align contacts on bo
t
om edge of pump with contacts inside
cradle and insert pump in cradle. Repeat for each additional
pump/cradle.
3. Charge
ba
t
ery completely (approximately 3 hours). The left
LED on the cradle will indicate charging status
(see Reading
Charge Status on Cradle LED)
.
Reading Charge Status on Cradle LED
The left LED on the charging cradle indicates ba