background image

Form 37714 Rev 2002

Page 4 of 13

www.skcinc.com

High Flow Applications (Cont)

5.  Cap the sample and send it with blanks and pertinent sampling information to a laboratory for analysis.

6.  Verify the flow.

a. Turn on the pump and reinstate the calibration train and sample medium.

b. Take three readings and record the average value as the post-sample flow rate. 

Do not adjust the pump flow rate during this step.

c. Compare the pre and post-sample flow rates. Note in sampling documentation if the values differ by more than ± 5%. 

Low Flow Applications (5 to 500 ml/min)

•  Single-tube sampling requires the All-in-One Low Flow Holder; see the operating instructions for the All-in-One for details on its operation. 

•  Multiple-tube sampling requires a Constant Pressure Controller (CPC) and a Dual, Tri, or Quad Adjustable Low Flow Tube Holder accessory; 

see the operating instructions for the CPC and Adjustable Low Flow Tube Holder for details on their operation.

Using the All-in-One Low Flow Holder (Figure 2)

Set/Calibrate Flow Rate for Single Tube

•  Allow pump to equilibrate after moving it from one temperature extreme to another.

•  Charge pump battery completely before calibration and sampling.

•  Calibrate/verify pump flow rate before and after each sampling operation using the tube holder and 

pump to be used for sampling.

•  Set the flow rate through the pump first and then calibrate the flow rate through the representative 

sorbent tube.

•  Two small inlet holes on the bottom of the built-in CPC of the All-in-One can become blocked. 

Periodically inspect and, if needed, clean with a small pick and blow particles away with a puff of air.

Pump Flow Rate

1.  Ensure that the battery is fully charged and that the pump has run for 5 minutes before 

calibrating. Leave the pump running.

2.  Prepare the calibrator 

(see calibrator instructions)

. Using flexible tubing, connect the calibrator outlet (suction port) to the pump inlet.

3. 

Set the pump flow rate to 1.5 L/min. Using a small screwdriver, turn the flow adjustment screw on top of the pump 

(Figure 1)

clockwise

to increase flow or 

counterclockwise

 to decrease flow, until the calibrator indicates 1.5 L/min.

4.  Remove tubing from the pump inlet.

Flow Rate Through Sorbent Tube

1.  Attach the Tygon tubing of the All-in-One to the pump inlet. 

Note: 

If the pump flow faults shortly after the holder is attached, check that 

the needle valve is open by using a small flat-head screwdriver to turn the flow adjust screw on the holder 

slightly

counterclockwise. If flow fault 

continues, check that the two small holes on the bottom of the built-in CPC are not blocked. If needed, clean them with a small pick and blow any 
particles away with a puff of air. 

2. 

Break tips off the representative sorbent tube and insert it into the rubber sleeve on the All-in-One

 (arrow on the tube points toward 

the holder).

Note:

 Ensure that the sorbent tube fits snugly in the rubber sleeve prior to sampling to avoid air leakage. Two sleeves, each with a 

different inner diameter, are supplied with the All-in-One.

3. 

Using flexible tubing, connect the calibrator outlet to the inlet of the representative sorbent tube in the holder. 

See calibration train above 

right

.

  In the next step, do not shut off flow completely with flow adjust screw or use an oversize screwdriver to 

adjust flow — valve or thread seat damage may result.

4. 

Do not adjust the flow on the pump.

 Adjust the flow rate through the representative sorbent tube 

using a small flat-head screwdriver to turn the flow adjust screw on the All-in-One 

(counterclockwise 

to increase, clockwise to decrease)

 until the calibrator displays the method-specified flow rate. Take a 

minimum of three readings and record the average flow rate, as per OSHA/NIOSH instructions.

5. 

When calibration is complete, turn off the pump by pressing the on/off button. Disconnect the calibrator 
and tubing from the representative sorbent tube inlet.

Turn screw on holder 

to adjust flow.

Flow 

adjust 

screw

Calibration train with All-in-One 

Low Flow Holder connected 

to AirChek 52

All-in-One

Tygon 
tubing

Pump 

inlet

Flow adjust screw 

(needle valve)

Suction 

port

Содержание 224-52

Страница 1: ...mpingers cyclones PPIs or filter cassettes Checking Pump Kit Contents Use the table below to verify that you received all items associated with the Cat No ordered If you are missing items contact SKC...

Страница 2: ...nd any warranty Ensure proper orientation of charging cable before plugging it into the charging jack Improper orientation contact will short circuit the battery and voids any warranty Short circuitin...

Страница 3: ...collection medium to the worker in the breathing zone and the pump to the worker s belt using the belt clip When using an impinger an in line trap SKC Cat No 225 22 or 225 22 01 is required between t...

Страница 4: ...ing a small screwdriver turn the flow adjustment screw on top of the pump Figure 1 clockwise to increase flow or counterclockwise to decrease flow until the calibrator indicates 1 5 L min 4 Remove tub...

Страница 5: ...o compensate due to excessive back pressure or a low battery condition exists it will shut down and timing functions will pause The LCD will display either a battery shaped icon or a flow fault icon d...

Страница 6: ...le tubing connect the calibrator outlet suction port to the pump inlet 3 Calculate the sum of all tube flow rates The maximum flow rate through any one tube is 500 ml min a If the sum is 1000 ml min s...

Страница 7: ...to decrease until the calibrator indicates the method specified flow rate Do not adjust the flow rate on the pump Note For tri and quad models first rotate each anti tamper cover to expose the flow ad...

Страница 8: ...elapsed run time Possible Displays During Sampling Flow or Battery Fault Shutdown If the pump is unable to compensate due to excessive back pressure or a low battery condition exists it will shut dow...

Страница 9: ...any warranty and the UL Listing for intrinsic safety 1 Loosen two security screws on bottom of pump case and allow them to fall to the out position 2 Pinch battery pack release arms and pull the botto...

Страница 10: ...4 26 CPC Protective Tube Covers for Sorbent Tubes Use with adjustable low flow tube holders listed above Type A 6 mm OD x 70 mm L included in Low Flow Kit above 224 29A Type B 8 mm OD x 110 mm L 224 2...

Страница 11: ...n illustration N A PC Board P79592 N A Screw and O ring Set complete for one pump P22402 External Case No Description Cat No 1 Battery Pack NiMH P78011AMH 2 Battery Pack Release Arms N A 3 Battery Pac...

Страница 12: ...mpensate for longer than 15 seconds due to excessive back pressure the pump enters flow fault During flow fault the pump stops the fault icon appears on the display and elapsed time remains on the dis...

Страница 13: ...5 48 41 3 0 42 38 Table 1 Pump Run Time in Hours with NiMH Battery Appendix Cont Typical Run Times Table 1 contains the typical run times achieved when using a fully charged NiMH battery pack Data is...

Отзывы: