background image

Form 37714 Rev 2002

Page 12 of 13

www.skcinc.com

Appendix

Performance Profile

Flow Range

1000 to 3000 ml/min 

5 to 500 ml/min requires low flow accessories. See Accessories.

Flow Control

Holds constant flow to ± 5% of set-point after calibration

Compensation Range 

(back pressure capability)

1000 ml/min up to 25 inches water back pressure
2000 ml/min up to 25 inches water back pressure
2500 ml/min up to 20 inches water back pressure
3000 ml/min up to 15 inches water back pressure

Typical Back Pressure of 
Sampling Media 

(inches water)

Flow Rate (L/min)

1.0

1.5

2.0

2.5

Filter/Pore Size (

μ

m)

25-mm MCE/0.8

6

9

12

15

25-mm MCE/0.45

14

22

28

35

37-mm MCE/0.8

2

3

4

5

37-mm PVC/5.0

1

1

2

2

Compare the information in this table to pump compensation range to determine appropriate applications.

Run Time

Nickel-metal hydride (NiMH) battery:

 12 hrs minimum at 2000 ml/min and up to 25 inches 

water back pressure; depends on media used. 

See Table 1.

Battery Eliminator:

 Pump provides extended runs.

Power Supply

Rechargeable 4.8-V NiMH battery pack,

 3.3-Ah capacity

Battery Eliminator

 is available 

(see Accessories)

; use voids the UL Listing for intrinsic safety.

Battery Charging Time

(varies with battery capacity
and level of discharge)

6 to 8.5 hrs with PowerFlex charger

Temperature

Operating:

 32 to 104 F (0 to 40 C)

Storage:

 -4 to 113 F (-20 to 45 C)

Charging:

 50 to 113 F (10 to 45 C)

Humidity

0 to 95% non-condensing

Flow Fault

If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump 
enters flow fault. During flow fault, the pump stops, the fault icon appears on the display, and elapsed 
time remains on the display. Auto-restart is attempted up to 5 times. 

Note:

 Elapsed time information 

will be lost on restart.

Low Battery Fault

LCD displays low battery icon and pump shuts down. LCD remains on.

Time Display

LCD shows elapsed pump run time in minutes up to 99,999. 

Note: 

Older pump models display elapsed time up to three digits (999) and then roll back to zero.

LCD Indicator

LCD always remains on. Displayed elapsed run time information resets when pump is started.

Noise Level

62.5 dBA* - pump without case
55 dBA* - pump housed in noise-reducing case 

(optional accessory Cat. No. 224-96C, see Accessories)

* Measured 3 ft (1 m) from pump operating at 2 L/min with a 37-mm, 0.8-

μ

m MCE filter cassette

RFI/EMI Shielding

 

27 to 1000 MHz

Intrinsic Safety

UL Listed for Intrinsic Safety: Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, 
Groups E, F, G; and, Class III, Temperature Code T3C. ATEX-approved models available.

 Contact SKC.

Dimensions

5 x 3 x 1.75 in (12.7 x 7.6 x 4.4 cm)

Weight

20 oz (567 grams)

Multiple-tube Sampling

 

Requires additional accessories to sample with up to four sorbent tubes simultaneously at flows from 
5 to 500 ml/min, dependent on back pressure. See Accessories.

Tubing

Requires 1/4-inch ID tubing

Certifications

•  Intrinsic safety: UL Class I, Groups A, B, C, and D; Class II, Groups E, F, and G; Class III hazardous 

locations when used with SKC battery pack model P78011AMH (SKC Cat. No. P78011AMH)

• CE marked
•  ATEX-approved models available

 

Содержание 224-52

Страница 1: ...mpingers cyclones PPIs or filter cassettes Checking Pump Kit Contents Use the table below to verify that you received all items associated with the Cat No ordered If you are missing items contact SKC...

Страница 2: ...nd any warranty Ensure proper orientation of charging cable before plugging it into the charging jack Improper orientation contact will short circuit the battery and voids any warranty Short circuitin...

Страница 3: ...collection medium to the worker in the breathing zone and the pump to the worker s belt using the belt clip When using an impinger an in line trap SKC Cat No 225 22 or 225 22 01 is required between t...

Страница 4: ...ing a small screwdriver turn the flow adjustment screw on top of the pump Figure 1 clockwise to increase flow or counterclockwise to decrease flow until the calibrator indicates 1 5 L min 4 Remove tub...

Страница 5: ...o compensate due to excessive back pressure or a low battery condition exists it will shut down and timing functions will pause The LCD will display either a battery shaped icon or a flow fault icon d...

Страница 6: ...le tubing connect the calibrator outlet suction port to the pump inlet 3 Calculate the sum of all tube flow rates The maximum flow rate through any one tube is 500 ml min a If the sum is 1000 ml min s...

Страница 7: ...to decrease until the calibrator indicates the method specified flow rate Do not adjust the flow rate on the pump Note For tri and quad models first rotate each anti tamper cover to expose the flow ad...

Страница 8: ...elapsed run time Possible Displays During Sampling Flow or Battery Fault Shutdown If the pump is unable to compensate due to excessive back pressure or a low battery condition exists it will shut dow...

Страница 9: ...any warranty and the UL Listing for intrinsic safety 1 Loosen two security screws on bottom of pump case and allow them to fall to the out position 2 Pinch battery pack release arms and pull the botto...

Страница 10: ...4 26 CPC Protective Tube Covers for Sorbent Tubes Use with adjustable low flow tube holders listed above Type A 6 mm OD x 70 mm L included in Low Flow Kit above 224 29A Type B 8 mm OD x 110 mm L 224 2...

Страница 11: ...n illustration N A PC Board P79592 N A Screw and O ring Set complete for one pump P22402 External Case No Description Cat No 1 Battery Pack NiMH P78011AMH 2 Battery Pack Release Arms N A 3 Battery Pac...

Страница 12: ...mpensate for longer than 15 seconds due to excessive back pressure the pump enters flow fault During flow fault the pump stops the fault icon appears on the display and elapsed time remains on the dis...

Страница 13: ...5 48 41 3 0 42 38 Table 1 Pump Run Time in Hours with NiMH Battery Appendix Cont Typical Run Times Table 1 contains the typical run times achieved when using a fully charged NiMH battery pack Data is...

Отзывы: