background image

Form 37714 Rev 2002

Page 3 of 13

www.skcinc.com

Operation 

High Flow Applications (1000 to 3000 ml/min)

Set/Calibrate Flow Rate

•  Allow pump to equilibrate after moving it from one temperature extreme to another.

•  Charge battery completely before calibration and sampling.

1. 

Ensure that the battery is fully charged and that the pump has run for five minutes before  

 

calibrating. Leave the pump running.

2.  Prepare the calibrator. 

See calibrator instructions

3.  Set up a calibration train: Using flexible tubing, connect the calibrator outlet (suction port) 

to the representative sample medium inlet. Using ¼-inch Tygon tubing, connect the sample 
medium outlet to the pump inlet.

 See right

.

4. 

Using a screwdriver, set the pump flow rate by turning the flow adjustment screw on top of the pump 

(Figure 1)

clockwise

 to increase 

or 

counterclockwise

 to decrease flow, until the calibrator indicates the method-specified flow rate. Take a minimum of three readings 

and record the average flow rate, as per OSHA/NIOSH instructions.

5.  Press the on/off button to turn off the pump. Disconnect the calibrator and flexible tubing from the representative sample medium 

and proceed to 

Set Up/Sample

.

Set Up/Sample

•  Allow pump to equilibrate after moving it from one temperature extreme to another.

•  Protect sample pump from weather when sampling outdoors.

•  Powering the pump with any device other than the approved battery pack voids the UL Listing for intrinsic safety.

• Do 

not

 use the pump with the Battery Eliminator in hazardous locations. UL Listing for intrinsic safety is not in effect 

during pump operation with Battery Eliminator. 

•  Use of any device other than the approved battery pack or Battery Eliminator to power the pump voids any warranty.

•  Charge battery completely before calibration and sampling.

1.  Replace the representative sample medium with new unexposed medium for sample collection.

2.  Place the sample medium where appropriate for sampling.

a. For 

personal sampling,

 clip the sample collection medium to the worker in the breathing zone and the 

pump to the worker’s belt using the belt clip.

  When using an impinger, an in-line trap (SKC Cat. No. 225-22 or 225-22-01) is required between the pump and 

impinger to prevent fumes from accidentally being drawn into the sampler. Failure to use an appropriate in-line 
trap during impinger sampling voids any warranty.

b. 

For sampling with an impinger, 

mount the single or dual impinger and trap (SKC Cat. No. 225-20-01 or 

225-20-02) to the face of the sampler using the accessory mounting screws 

(Figure 1)

 or place them in a 

holster at the worker’s waist.

3.  Press the on/off button to clear the elapsed time displayed on the LCD and to start sampling.  Record the 

start time and other pertinent information. The LCD will automatically display elapsed pump run time. 

See below.

Possible Displays During Sampling

Flow or Battery Fault Shutdown -

 If the pump is unable to compensate due to excessive back pressure or a low battery 

condition exists, it will shut down and timing functions will pause.  The LCD will display either a battery-shaped icon 
or a flow fault icon (

) depending on the cause of the shutdown. Upon flow fault, the pump will attempt to restart up 

to five times.  To restart from flow fault, correct the blockage and press the on/off button twice. 

The elapsed run time 

display will reset to 0 when the pump is restarted.

 If the battery icon is displayed, recharge the battery before sampling. 

Displayed Elapsed Run Time - 

Elapsed run time is displayed continuously on the LCD. For elapsed times after 999 

minutes, the display still shows elapsed time but alternates between displaying the first two digits of the elapsed time 
and the last three digits. For example, a pump that has run for 1,440 minutes would display first “01” and then “440.” 
The display maximum is 99 999, which is 99,999 minutes.

4.  At the end of the sampling period, press the on/off button and record stop time and other pertinent information.

chek-mate 

Calibrator

Sample medium

Tygon 
tubing

Calibration train with filter cassette

Pump 

inlet

Suction 

port

Clip sample medium 

to worker and pump 

to belt.

Impinger holder on pump 

with impinger and trap

Содержание 224-52

Страница 1: ...mpingers cyclones PPIs or filter cassettes Checking Pump Kit Contents Use the table below to verify that you received all items associated with the Cat No ordered If you are missing items contact SKC...

Страница 2: ...nd any warranty Ensure proper orientation of charging cable before plugging it into the charging jack Improper orientation contact will short circuit the battery and voids any warranty Short circuitin...

Страница 3: ...collection medium to the worker in the breathing zone and the pump to the worker s belt using the belt clip When using an impinger an in line trap SKC Cat No 225 22 or 225 22 01 is required between t...

Страница 4: ...ing a small screwdriver turn the flow adjustment screw on top of the pump Figure 1 clockwise to increase flow or counterclockwise to decrease flow until the calibrator indicates 1 5 L min 4 Remove tub...

Страница 5: ...o compensate due to excessive back pressure or a low battery condition exists it will shut down and timing functions will pause The LCD will display either a battery shaped icon or a flow fault icon d...

Страница 6: ...le tubing connect the calibrator outlet suction port to the pump inlet 3 Calculate the sum of all tube flow rates The maximum flow rate through any one tube is 500 ml min a If the sum is 1000 ml min s...

Страница 7: ...to decrease until the calibrator indicates the method specified flow rate Do not adjust the flow rate on the pump Note For tri and quad models first rotate each anti tamper cover to expose the flow ad...

Страница 8: ...elapsed run time Possible Displays During Sampling Flow or Battery Fault Shutdown If the pump is unable to compensate due to excessive back pressure or a low battery condition exists it will shut dow...

Страница 9: ...any warranty and the UL Listing for intrinsic safety 1 Loosen two security screws on bottom of pump case and allow them to fall to the out position 2 Pinch battery pack release arms and pull the botto...

Страница 10: ...4 26 CPC Protective Tube Covers for Sorbent Tubes Use with adjustable low flow tube holders listed above Type A 6 mm OD x 70 mm L included in Low Flow Kit above 224 29A Type B 8 mm OD x 110 mm L 224 2...

Страница 11: ...n illustration N A PC Board P79592 N A Screw and O ring Set complete for one pump P22402 External Case No Description Cat No 1 Battery Pack NiMH P78011AMH 2 Battery Pack Release Arms N A 3 Battery Pac...

Страница 12: ...mpensate for longer than 15 seconds due to excessive back pressure the pump enters flow fault During flow fault the pump stops the fault icon appears on the display and elapsed time remains on the dis...

Страница 13: ...5 48 41 3 0 42 38 Table 1 Pump Run Time in Hours with NiMH Battery Appendix Cont Typical Run Times Table 1 contains the typical run times achieved when using a fully charged NiMH battery pack Data is...

Отзывы: